Feasibility of a novel mapping system combined with remote magnetic navigation for catheter ablation of premature ventricular contractions.


ABSTRACT:

Background

Remote magnetic navigation (RMN) is often used in combination with a 3-dimensional mapping system to perform catheter ablations. This study aim to investigate the feasibility and effectiveness of a novel 3D-mapping system, EnSite Precision, combined with RMN for catheter ablation of premature ventricular contractions (PVCs), and compared it to the procedures performed by CARTO3 with RMN.

Methods

Forty-three consecutive PVC patients were either ablated with the guidance of EnSite Precision (n = 22) or CARTO (n = 21) navigated by RMN. Procedure-related details, acute and long-term success were assessed.

Results

Patient characteristics between both the groups were similar (age: 47.1 ± 19.8 vs 47.1 ± 12.7, female: 63.6% vs 57.1%). No significant difference was found in the procedure time (99.5 ± 30.4 vs 92.9 ± 24.8 min, = 0.436), mapping time (18.6 ± 12.8 vs 15.5 ± 10.2 min, = 0.390), radiofrequency ablation time (333.4 ± 267.0 vs 469.3 ± 343.1 s, = 0.154), fluoroscopy time (4.0 ± 1.9 vs 3.8 ± 2.0 min, = 0.635), and X-ray dose (1.8 ± 1.4 vs 2.0 ± 1.2 Gycm2, = 0.649) between the two groups. No significant procedural complication occurred in either group. In addition, there was no significant differences regarding the acute success rate (90.9% vs 90.5%, = 0.961) and long-term success rate (86.4% vs 81.0%, = 0.631) after 16.2 ± 6.2 months of follow-up between the two groups.

Conclusions

RMN combined with EnSite Precision mapping system is effective and safe for catheter ablation of PVCs.

SUBMITTER: Dang S 

PROVIDER: S-EPMC6457387 | biostudies-literature | 2019 Apr

REPOSITORIES: biostudies-literature

altmetric image

Publications

Feasibility of a novel mapping system combined with remote magnetic navigation for catheter ablation of premature ventricular contractions.

Dang Shipeng S   Jons Christian C   Jacobsen Peter Karl PK   Pehrson Steen S   Chen Xu X  

Journal of arrhythmia 20190131 2


<h4>Background</h4>Remote magnetic navigation (RMN) is often used in combination with a 3-dimensional mapping system to perform catheter ablations. This study aim to investigate the feasibility and effectiveness of a novel 3D-mapping system, EnSite Precision, combined with RMN for catheter ablation of premature ventricular contractions (PVCs), and compared it to the procedures performed by CARTO3 with RMN.<h4>Methods</h4>Forty-three consecutive PVC patients were either ablated with the guidance  ...[more]

Similar Datasets

| S-EPMC8126685 | biostudies-literature
| S-EPMC8666643 | biostudies-literature
| S-EPMC5437699 | biostudies-literature
| S-EPMC4672078 | biostudies-literature
| S-EPMC6457378 | biostudies-literature
| S-EPMC6601165 | biostudies-literature
| S-EPMC9151449 | biostudies-literature
| S-EPMC6916391 | biostudies-literature