Motion robust high resolution 3D free-breathing pulmonary MRI using dynamic 3D image self-navigator.


ABSTRACT: PURPOSE:To achieve motion robust high resolution 3D free-breathing pulmonary MRI utilizing a novel dynamic 3D image navigator derived directly from imaging data. METHODS:Five-minute free-breathing scans were acquired with a 3D ultrashort echo time (UTE) sequence with 1.25 mm isotropic resolution. From this data, dynamic 3D self-navigating images were reconstructed under locally low rank (LLR) constraints and used for motion compensation with one of two methods: a soft-gating technique to penalize the respiratory motion induced data inconsistency, and a respiratory motion-resolved technique to provide images of all respiratory motion states. RESULTS:Respiratory motion estimation derived from the proposed dynamic 3D self-navigator of 7.5 mm isotropic reconstruction resolution and a temporal resolution of 300 ms was successful for estimating complex respiratory motion patterns. This estimation improved image quality compared to respiratory belt and DC-based navigators. Respiratory motion compensation with soft-gating and respiratory motion-resolved techniques provided good image quality from highly undersampled data in volunteers and clinical patients. CONCLUSION:An optimized 3D UTE sequence combined with the proposed reconstruction methods can provide high-resolution motion robust pulmonary MRI. Feasibility was shown in patients who had irregular breathing patterns in which our approach could depict clinically relevant pulmonary pathologies. Magn Reson Med 79:2954-2967, 2018. © 2017 International Society for Magnetic Resonance in Medicine.

SUBMITTER: Jiang W 

PROVIDER: S-EPMC6474413 | biostudies-literature | 2018 Jun

REPOSITORIES: biostudies-literature

altmetric image

Publications

Motion robust high resolution 3D free-breathing pulmonary MRI using dynamic 3D image self-navigator.

Jiang Wenwen W   Ong Frank F   Johnson Kevin M KM   Nagle Scott K SK   Hope Thomas A TA   Lustig Michael M   Larson Peder E Z PEZ  

Magnetic resonance in medicine 20171011 6


<h4>Purpose</h4>To achieve motion robust high resolution 3D free-breathing pulmonary MRI utilizing a novel dynamic 3D image navigator derived directly from imaging data.<h4>Methods</h4>Five-minute free-breathing scans were acquired with a 3D ultrashort echo time (UTE) sequence with 1.25 mm isotropic resolution. From this data, dynamic 3D self-navigating images were reconstructed under locally low rank (LLR) constraints and used for motion compensation with one of two methods: a soft-gating techn  ...[more]

Similar Datasets

| S-EPMC7806246 | biostudies-literature
| S-EPMC4065644 | biostudies-literature
| S-EPMC7341067 | biostudies-literature
| S-EPMC6949392 | biostudies-literature
| S-EPMC7904595 | biostudies-literature
| S-EPMC6699991 | biostudies-literature
| S-EPMC9246907 | biostudies-literature
| S-EPMC9785439 | biostudies-literature
| S-EPMC9298075 | biostudies-literature
| S-EPMC8858033 | biostudies-literature