ABSTRACT: Objective
To investigate the repeatability of choroidal thickness measurements determined from enhanced depth imaging optical coherence tomography (EDI-OCT) images of eyes after wearing single-vision spectacles (SV) and orthokeratology (ortho-k) lenses.Methods and analysis
Two EDI-OCT images of 40 children (SV: 20, ortho-k: 20) taken at a single visit were analysed twice. Subjects in the ortho-k group had been wearing ortho-k for 1-4 weeks. The choroidal thickness was determined from each image using a graph theory-based software and, where appropriate, manual correction of choroidal boundaries was undertaken by an experienced examiner.Results
The mean (±SD) choroidal thickness was 227.3±42.2 µm for the SV subjects and 251.1±54.4 µm for the ortho-k subjects. The interimage differences in choroidal thickness were -0.99±3.54 and -1.14±5.03 µm for the SV and ortho-k subjects, respectively, and the limits of agreement were +5.96 to -7.93 and +8.72 to -11.00 µm, respectively.Conclusion
The coefficients of repeatability of choroidal thickness measurements from two EDI-OCT images taken at a single visit were 7.08 µm (SV) and 10.06 µm (ortho-k), suggesting that a change in choroidal thickness of less than 10 µm may not indicate a real change resulting from ortho-k lens wear.
SUBMITTER: Lau JK
PROVIDER: S-EPMC6528750 | biostudies-literature | 2019
REPOSITORIES: biostudies-literature
Lau Jason K JK Cheung Sin Wan SW Collins Michael J MJ Cho Pauline P
BMJ open ophthalmology 20190423 1
<h4>Objective</h4>To investigate the repeatability of choroidal thickness measurements determined from enhanced depth imaging optical coherence tomography (EDI-OCT) images of eyes after wearing single-vision spectacles (SV) and orthokeratology (ortho-k) lenses.<h4>Methods and analysis</h4>Two EDI-OCT images of 40 children (SV: 20, ortho-k: 20) taken at a single visit were analysed twice. Subjects in the ortho-k group had been wearing ortho-k for 1-4 weeks. The choroidal thickness was determined ...[more]