Unknown

Dataset Information

0

SAMBADENA Hyperpolarization of 13C-Succinate in an MRI: Singlet-Triplet Mixing Causes Polarization Loss.


ABSTRACT: The signal enhancement provided by the hyperpolarization of nuclear spins of biological molecules is a highly promising technique for diagnostic imaging. To date, most 13C-contrast agents had to be polarized in an extra, complex or cost intensive polarizer. Recently, the in?situ hyperpolarization of a 13C contrast agent to >20?% was demonstrated without a polarizer but within the bore of an MRI system. This approach addresses some of the challenges of MRI with hyperpolarized tracers, i.?e. elevated cost, long production times, and loss of polarization during transfer to the detection site. Here, we demonstrate the first hyperpolarization of a biomolecule in aqueous solution in the bore of an MRI at field strength of 7?T within seconds. The 13C nucleus of 1-13C, 2,3-2H2-succinate was polarized to 11?% corresponding to a signal enhancement of approximately 18.000. Interesting effects during the process of the hydrogenation reaction which lead to a significant loss of polarization have been observed.

SUBMITTER: Berner S 

PROVIDER: S-EPMC6587320 | biostudies-literature | 2019 Jun

REPOSITORIES: biostudies-literature

altmetric image

Publications

SAMBADENA Hyperpolarization of <sup>13</sup>C-Succinate in an MRI: Singlet-Triplet Mixing Causes Polarization Loss.

Berner Stephan S   Schmidt Andreas B AB   Zimmermann Mirko M   Pravdivtsev Andrey N AN   Glöggler Stefan S   Hennig Jürgen J   von Elverfeldt Dominik D   Hövener Jan-Bernd JB  

ChemistryOpen 20190604 6


The signal enhancement provided by the hyperpolarization of nuclear spins of biological molecules is a highly promising technique for diagnostic imaging. To date, most <sup>13</sup>C-contrast agents had to be polarized in an extra, complex or cost intensive polarizer. Recently, the <i>in situ</i> hyperpolarization of a <sup>13</sup>C contrast agent to >20 % was demonstrated without a polarizer but within the bore of an MRI system. This approach addresses some of the challenges of MRI with hyperp  ...[more]

Similar Datasets

| S-EPMC5959305 | biostudies-literature
| S-EPMC6754226 | biostudies-literature
| S-EPMC5510972 | biostudies-literature
| S-EPMC6040279 | biostudies-literature