Artifacts in pulse transit time measurements using standard patient monitoring equipment.


ABSTRACT:

Objective

Pulse transit time (PTT) refers to the time it takes a pulse wave to travel between two arterial sites. PTT can be estimated, amongst others, using the electrocardiogram (ECG) and photoplethysmogram (PPG). Because we observed a sawtooth artifact in the PTT while using standard patient monitoring equipment for ECG and PPG, we explored the reasons for this artifact.

Methods

PPG and ECG were simulated at a heartrate of both 100 and 160 beats per minute while using a Masimo PPG post-processing module and a Philips patient monitor setup at the neonatal intensive care unit. Two different post-processing modules were used. PTT was defined as the difference between the R-peak in the ECG and the point of 50% increase in the PPG.

Results

A sawtooth artifact was seen in all simulations. Both length (59.2 to 72.4 s) and amplitude (30.8 to 36.0 ms) of the sawtooth were dependent on the post-processing module used. Furthermore, the absolute PTT value differed up to 250 ms depending on post-processing module and heart rate. The sawtooth occurred because the PPG wave continuously showed a minimal prolongation during the length of the sawtooth, followed by a sudden shortening. Both artifacts were generated in the post-processing module containing Masimo algorithms.

Conclusion

Post-processing of the PPG signal in the Masimo module of the Philips patient monitor introduces a sawtooth in PPG and derived PTT. This sawtooth, together with a large module-dependent absolute difference in PTT, renders the thus-derived PTT insufficient for clinical purposes.

SUBMITTER: Bennis FC 

PROVIDER: S-EPMC6588249 | biostudies-literature | 2019

REPOSITORIES: biostudies-literature

altmetric image

Publications

Artifacts in pulse transit time measurements using standard patient monitoring equipment.

Bennis Frank C FC   van Pul Carola C   van den Bogaart Jarno J L JJL   Andriessen Peter P   Kramer Boris W BW   Delhaas Tammo T  

PloS one 20190621 6


<h4>Objective</h4>Pulse transit time (PTT) refers to the time it takes a pulse wave to travel between two arterial sites. PTT can be estimated, amongst others, using the electrocardiogram (ECG) and photoplethysmogram (PPG). Because we observed a sawtooth artifact in the PTT while using standard patient monitoring equipment for ECG and PPG, we explored the reasons for this artifact.<h4>Methods</h4>PPG and ECG were simulated at a heartrate of both 100 and 160 beats per minute while using a Masimo  ...[more]

Similar Datasets

| S-EPMC7197410 | biostudies-literature
| S-EPMC9319785 | biostudies-literature
| S-EPMC6733479 | biostudies-literature
| S-EPMC6711549 | biostudies-literature
| S-EPMC7039871 | biostudies-literature
| S-EPMC5601224 | biostudies-literature
| S-EPMC7544103 | biostudies-literature
| S-EPMC6179208 | biostudies-literature
| S-EPMC7002624 | biostudies-literature
| S-EPMC6418715 | biostudies-literature