Unknown

Dataset Information

0

Impact of Leaflet Laceration on Transcatheter Aortic Valve-in-Valve Washout: BASILICA to Solve Neosinus and Sinus Stasis.


ABSTRACT: OBJECTIVES:The aim of this study was to evaluate any potential leaflet washout benefits after bioprosthetic or native aortic scallop intentional laceration to prevent iatrogenic coronary artery obstruction during TAVR (BASILICA) in transcatheter valve-in-valve (ViV) in the context of leaflet thrombosis. BACKGROUND:Leaflet thrombosis after transcatheter aortic valve replacement is secondary to flow stasis in both the sinus and neosinus. Strategies to improve washout and ameliorate neosinus and sinus flow velocities may have the potential to mitigate the occurrence of clinical and subclinical leaflet thrombosis. METHODS:A 23-mm Edwards SAPIEN 3 and a 26-mm Medtronic Evolut were deployed in a 23-mm transparent surgical aortic valve model before and after leaflet laceration. The valves were placed in the aortic position of a pulse duplicator flow loop. Particle image velocimetry was performed to quantify sinus flow hemodynamic status. A tracing fluorescent dye was injected to evaluate the number of cycles to washout in both regions of interest. RESULTS:The leaflet laceration procedure led to an increase in the velocities in the sinus and the neosinus by 50% for Evolut ViV and 61.9% for SAPIEN 3 ViV. In addition, leaflet laceration led to a reduction in overall cycles to washout in the neosinus by at least 56% with the Evolut and 54.5% with the SAPIEN 3 and in the sinus by at least 16.7% with the Evolut and 60.8% with the SAPIEN. CONCLUSIONS:Leaflet laceration using a BASILICA-type approach may hold the potential to mitigate neosinus and sinus flow stasis. Controlled in vivo trials are necessary to establish the potential benefit of BASILICA to reduce the occurrence of leaflet thrombosis.

SUBMITTER: Hatoum H 

PROVIDER: S-EPMC6613808 | biostudies-literature | 2019 Jul

REPOSITORIES: biostudies-literature

altmetric image

Publications

Impact of Leaflet Laceration on Transcatheter Aortic Valve-in-Valve Washout: BASILICA to Solve Neosinus and Sinus Stasis.

Hatoum Hoda H   Maureira Pablo P   Lilly Scott S   Dasi Lakshmi Prasad LP  

JACC. Cardiovascular interventions 20190701 13


<h4>Objectives</h4>The aim of this study was to evaluate any potential leaflet washout benefits after bioprosthetic or native aortic scallop intentional laceration to prevent iatrogenic coronary artery obstruction during TAVR (BASILICA) in transcatheter valve-in-valve (ViV) in the context of leaflet thrombosis.<h4>Background</h4>Leaflet thrombosis after transcatheter aortic valve replacement is secondary to flow stasis in both the sinus and neosinus. Strategies to improve washout and ameliorate  ...[more]

Similar Datasets

| S-EPMC5479725 | biostudies-literature
| S-EPMC11308060 | biostudies-literature
| S-EPMC6664295 | biostudies-literature
| S-EPMC7047047 | biostudies-literature
| S-EPMC7501921 | biostudies-literature
| S-EPMC6724191 | biostudies-literature
| S-EPMC8302883 | biostudies-literature
| S-EPMC8549999 | biostudies-literature
| S-EPMC6309616 | biostudies-literature
| S-EPMC8311375 | biostudies-literature