Texture analysis in 177Lu SPECT phantom images: Statistical assessment of uniformity requirements using texture features.


ABSTRACT: The purpose of this study was to apply texture analysis (TA) to evaluate the uniformity of SPECT images reconstructed with the 3D Ordered Subsets Expectation Maximization (3D-OSEM) algorithm. For this purpose, a cylindrical homogeneous phantom filled with 177Lu was used and a total of 24 spherical volumes of interest (VOIs) were considered inside the phantom. The location of the VOIs was chosen in order to define two different configurations, i.e. gravity and radial configuration. The former configuration was used to investigate the uniformity of distribution of 177Lu inside the phantom, while the latter configuration was used to investigate the lack of uniformity from center towards edge of the images. For each VOI, the trend of different texture features considered as a function of 3D-OS

SUBMITTER: Sarnelli A 

PROVIDER: S-EPMC6668785 | biostudies-literature | 2019