Review on cochlear implant electrode array tip fold-over and scalar deviation.


ABSTRACT:

Objective

Determine the occurrence rate of cochlear implant (CI) electrode tip fold-over and electrode scalar deviation as reported in patient cases with different commercial electrode types.

Data-sources

PubMed search for identifying peer-reviewed articles published till 2018 on CI electrode tip fold-over and scalar deviation. Key-words for searching were "Cochlear electrode tip fold-over", "Cochlear electrode scalar position" and "Cochlear electrode scalar location".

Articles-selection

Only if electrode related issues were investigated in patient cases. 38 articles met the inclusion-criteria.

Results

13 articles on electrode tip fold-over issue covering 3177 implanted ears, out of which 50 ears were identified with electrode tip fold-over with an occurrence rate of 1.57%. Out of 50 ears, 43 were implanted with pre-curved electrodes and the remaining 7 with lateral-wall electrodes. One article reported on both tip fold-over and scalar deviation. 26 articles reported on the electrode scalar deviation covering an overall number of 2046 ears out of which, 458 were identified with electrode scalar deviation at a rate of 22.38%. After removing the studies that did not report on the number of electrodes per electrode type, it was 1324 ears implanted with pre-curved electrode and 507 ears with lateral-wall electrode. Out of 1324 pre-curved electrode implanted ears, 424 were reported with scalar deviation making an occurrence rate of 32%. Out of 507 lateral-wall electrode implanted ears, 43 were associated with scalar deviation at an occurrence rate of 6.7%.

Conclusion

This literature review revealing the fact of higher rate of electrode insertion trauma associated with pre-curved electrode type irrespective of CI brand is one step closer to obsolete it from the clinical practice in the interest of patient's cochlear health.

SUBMITTER: Dhanasingh A 

PROVIDER: S-EPMC6712287 | biostudies-literature | 2019 Sep

REPOSITORIES: biostudies-literature

altmetric image

Publications

Review on cochlear implant electrode array tip fold-over and scalar deviation.

Dhanasingh Anandhan A   Jolly Claude C  

Journal of otology 20190109 3


<h4>Objective</h4>Determine the occurrence rate of cochlear implant (CI) electrode tip fold-over and electrode scalar deviation as reported in patient cases with different commercial electrode types.<h4>Data-sources</h4>PubMed search for identifying peer-reviewed articles published till 2018 on CI electrode tip fold-over and scalar deviation. Key-words for searching were "Cochlear electrode tip fold-over", "Cochlear electrode scalar position" and "Cochlear electrode scalar location".<h4>Articles  ...[more]

Similar Datasets

| S-EPMC3443690 | biostudies-literature
| S-EPMC6549524 | biostudies-literature
| S-EPMC5335919 | biostudies-literature
| S-EPMC8990937 | biostudies-literature
| S-EPMC2538331 | biostudies-other
| S-EPMC8246990 | biostudies-literature
| S-EPMC9271767 | biostudies-literature
| S-EPMC10583924 | biostudies-literature
| S-EPMC6641467 | biostudies-literature
| S-EPMC7039896 | biostudies-literature