Unknown

Dataset Information

0

Gold Plate Electrodes Functionalized by Multiwall Carbon Nanotube Film for Potentiometric Thallium(I) Detection.


ABSTRACT: Solid-contact potentiometric ion-selective electrodes (SC-ISEs) for thallium determination have been designed using multiwall carbon nanotubes (MWCNTs) as the ion-to-electron transducer. Dispersed MWCNTs were drop-casted over a gold plate electrode. Two different crown ethers were used in the sensing membrane for the recognition of thallium (I). Sensorsbased on dibenzo-18-crown-6 (DB18C6) as a neutral carrier and NaTPB as an anionic additive exhibited a near Nernstian response of 57.3 mV/decade towards Tl+ ions over the activity range 4.5 × 10-6-7.0 × 10-4 M, with a limit of detection of 3.2 × 10-7 M. The time required to achieve 95% of the steadyequilibrium potential was <10 s. The complex formation constant (log ?ML) between dibenzo-18-crown-6 and thallium (I) (i.e., 5.99) was measured using the sandwich membrane technique. The potential response was pH independent over the range 3.0-9.5. The introduction of MWCNTs as an electron-ion-transducer layer between gold plate and the sensing membrane lead to a smaller membrane resistance and a large double layer capacitance, which was proven using impedance spectra and chronopotentiometry (i.e., 114.9 ± 12 k?, 52.1 ± 3.3 pF, 200 ± 13.2 k?, and 50 ± 4.2 µF). Additionally, reduction ofthe water layer between the sensing membrane and the underlying conductor wastested. Thus, it is clear that MWCNTs can be used as a transducing layer in SC-ISEs. The proposed sensor was introduced as an indicator electrode for potentiometric titration of single and ternary mixtures of I-, Br-, and S2- anions.

SUBMITTER: Hassan SSM 

PROVIDER: S-EPMC6723907 | biostudies-literature | 2019 Aug

REPOSITORIES: biostudies-literature

altmetric image

Publications

Gold Plate Electrodes Functionalized by Multiwall Carbon Nanotube Film for Potentiometric Thallium(I) Detection.

Hassan Saad S M SSM   Abdelbasir Sabah M SM   Fathy M Abdelwahab MA   Amr Abd El-Galil E AEE   Al-Omar Mohamed A MA   Kamel Ayman H AH  

Nanomaterials (Basel, Switzerland) 20190814 8


Solid-contact potentiometric ion-selective electrodes (SC-ISEs) for thallium determination have been designed using multiwall carbon nanotubes (MWCNTs) as the ion-to-electron transducer. Dispersed MWCNTs were drop-casted over a gold plate electrode. Two different crown ethers were used in the sensing membrane for the recognition of thallium (I). Sensorsbased on dibenzo-18-crown-6 (DB18C6) as a neutral carrier and NaTPB as an anionic additive exhibited a near Nernstian response of 57.3 mV/decade  ...[more]

Similar Datasets

2021-12-29 | GSE172278 | GEO
| S-EPMC8153315 | biostudies-literature
| S-EPMC6566365 | biostudies-literature
| S-EPMC7830566 | biostudies-literature
| S-EPMC7399653 | biostudies-literature
| S-EPMC4985658 | biostudies-literature
| S-EPMC5056342 | biostudies-literature
| PRJNA722743 | ENA
| S-EPMC5736787 | biostudies-literature
| S-EPMC8314270 | biostudies-literature