Ontology highlight
ABSTRACT: Background
Sildenafil and calcitonin gene-related peptide both dilate the intradural segments of the middle meningeal artery measured with 3.0 tesla (T) MR angiography. Here we hypothesized that an increase in field strength to 7.0 T and concomitant enhanced voxel resolution would lower variance in measurements of dilation in the intradural middle meningeal artery.Methods
Five subjects completed two sessions at respectively 3.0 T and 7.0 T. Each session comprised MR angiography scans once before and twice after administration of sildenafil, calcitonin gene-related peptide or placebo in a three-way, crossover, double-blind, placebo-controlled design.Results
Standard deviations of arterial circumference revealed no difference between 3.0 T and 7.0 T measurements (p = 0.379). We found a decrease in standard deviation from our original angiography analysis software (QMra) to a newer (LAVA) software package (p < 0.001). Furthermore, we found that the dilation after sildenafil and calcitonin gene-related peptide were comparable between 3.0 T and 7.0 T.Conclusions
Our findings suggest no gain from the increase in voxel resolution but cemented dilatory findings from earlier. The implemented software update improved variance in circumference measurements in the intradural middle meningeal artery, which should be exploited in future studies.Trial registration
The study is part of a parent study, which is registered at ClinicalTrials.gov ( NCT03143465 ).
SUBMITTER: Christensen CE
PROVIDER: S-EPMC6734314 | biostudies-literature | 2019 May
REPOSITORIES: biostudies-literature
Christensen Casper Emil CE Younis Samaira S Lindberg Ulrich U Boer Vincent Oltman VO de Koning Patrick P Petersen Esben Thade ET Paulson Olaf Bjarne OB Larsson Henrik Bo Wiberg HBW Amin Faisal Mohammad FM Ashina Messoud M
The journal of headache and pain 20190506 1
<h4>Background</h4>Sildenafil and calcitonin gene-related peptide both dilate the intradural segments of the middle meningeal artery measured with 3.0 tesla (T) MR angiography. Here we hypothesized that an increase in field strength to 7.0 T and concomitant enhanced voxel resolution would lower variance in measurements of dilation in the intradural middle meningeal artery.<h4>Methods</h4>Five subjects completed two sessions at respectively 3.0 T and 7.0 T. Each session comprised MR angiography s ...[more]