Fat navigators and Moire phase tracking comparison for motion estimation and retrospective correction.


ABSTRACT:

Purpose

To compare motion tracking by two modern methods (fat navigators [FatNavs] and Moiré phase tracking [MPT]) as well as their performance for retrospective correction of very high resolution acquisitions.

Methods

A direct comparison of FatNavs and MPT motion parameters was performed for several deliberate motion patterns to estimate the agreement between methods. In addition, two different navigator resolution were applied. 0.5 mm isotropic MP2RAGE images with simultaneous MPT and FatNavs tracking were acquired in 9 cooperative subjects with no intentional motion. Retrospective motion corrections based on both tracking modalities were compared qualitatively and quantitatively. The FatNavs impact on quantitative T1 maps was also investigated.

Results

Both methods showed good agreement within a 0.3 mm/° margin in subjects that moved very little. Higher resolution FatNavs (2 mm) showed overall better agreement with MPT than 4 mm resolution ones, except for fast and large motion. The retrospective motion corrections based on MPT or FatNavs were at par in 33 cases out of 36, and visibly improved image quality compared to the uncorrected images. In separate fringe cases, both methods suffered from their respective potential shortcomings: unreliable marker attachment for MPT and poor temporal resolution for FatNavs. The magnetization transfer induced by the navigator RF pulses had a visible impact on the T1 values distribution, with a shift of the gray and white matter peaks of 12 ms at most.

Conclusion

This work confirms both FatNavs and MPT as excellent retrospective motion correction methods for very high resolution imaging of cooperative subjects.

SUBMITTER: Gretsch F 

PROVIDER: S-EPMC6778708 | biostudies-literature | 2020 Jan

REPOSITORIES: biostudies-literature

altmetric image

Publications

Fat navigators and Moiré phase tracking comparison for motion estimation and retrospective correction.

Gretsch Frédéric F   Mattern Hendrik H   Gallichan Daniel D   Speck Oliver O  

Magnetic resonance in medicine 20190809 1


<h4>Purpose</h4>To compare motion tracking by two modern methods (fat navigators [FatNavs] and Moiré phase tracking [MPT]) as well as their performance for retrospective correction of very high resolution acquisitions.<h4>Methods</h4>A direct comparison of FatNavs and MPT motion parameters was performed for several deliberate motion patterns to estimate the agreement between methods. In addition, two different navigator resolution were applied. 0.5 mm isotropic MP2RAGE images with simultaneous M  ...[more]

Similar Datasets

| S-EPMC6945502 | biostudies-literature
| S-EPMC6258267 | biostudies-literature
| S-EPMC7839562 | biostudies-literature
| S-EPMC6626557 | biostudies-literature
| S-EPMC3320676 | biostudies-other
| S-EPMC3539196 | biostudies-other
| S-EPMC6593735 | biostudies-literature
| S-EPMC4754677 | biostudies-literature
| S-EPMC6170141 | biostudies-literature
| S-EPMC6191114 | biostudies-literature