Unknown

Dataset Information

0

Remote heart rate monitoring - Assessment of the Facereader rPPg by Noldus.


ABSTRACT: Remote photoplethysmography (rPPG) allows contactless monitoring of human cardiac activity through a video camera. In this study, we assessed the accuracy and precision for heart rate measurements of the only consumer product available on the market, namely the FacereaderTM rPPG by Noldus, with respect to a gold standard electrocardiograph. Twenty-four healthy participants were asked to sit in front of a computer screen and alternate two periods of rest with two stress tests (i.e. Go/No-Go task), while their heart rate was simultaneously acquired for 20 minutes using the ECG criterion measure and the FacereaderTM rPPG. Results show that the FacereaderTM rPPG tends to overestimate lower heart rates and underestimate higher heart rates compared to the ECG. The Facereader™ rPPG revealed a mean bias of 9.8 bpm, the 95% limits of agreement (LoA) ranged from almost -30 up to +50 bpm. These results suggest that whilst the rPPG FacereaderTM technology has potential for contactless heart rate monitoring, its predictions are inaccurate for higher heart rates, with unacceptable precision across the entire range, rendering its estimates unreliable for monitoring individuals.

SUBMITTER: Benedetto S 

PROVIDER: S-EPMC6874325 | biostudies-literature | 2019

REPOSITORIES: biostudies-literature

altmetric image

Publications

Remote heart rate monitoring - Assessment of the Facereader rPPg by Noldus.

Benedetto Simone S   Caldato Christian C   Greenwood Darren C DC   Bartoli Nicola N   Pensabene Virginia V   Actis Paolo P  

PloS one 20191122 11


Remote photoplethysmography (rPPG) allows contactless monitoring of human cardiac activity through a video camera. In this study, we assessed the accuracy and precision for heart rate measurements of the only consumer product available on the market, namely the FacereaderTM rPPG by Noldus, with respect to a gold standard electrocardiograph. Twenty-four healthy participants were asked to sit in front of a computer screen and alternate two periods of rest with two stress tests (i.e. Go/No-Go task)  ...[more]

Similar Datasets

| S-EPMC6734354 | biostudies-literature
| S-EPMC5831032 | biostudies-literature
| S-EPMC3600881 | biostudies-other
| S-EPMC8856830 | biostudies-literature
| S-EPMC7819664 | biostudies-literature
| S-EPMC8102149 | biostudies-literature
| S-EPMC10573601 | biostudies-literature
| S-EPMC8160717 | biostudies-literature
| S-EPMC10486103 | biostudies-literature
| S-EPMC9882593 | biostudies-literature