Bismuth shield affecting CT image quality and radiation dose in adjacent and distant zones relative to shielding surface: A phantom study.


ABSTRACT: BACKGROUND:To quantify image quality and radiation doses in regions adjacent to and distant from bismuth shields in computed tomography (CT). METHODS:An American College of Radiology accreditation phantom with four solid rods embedded in a water-like background was scanned to verify CT number (CTN) accuracy when using bismuth shields. CTNs, image noise, and contrast-to-noise ratios (CNRs) were determined in the phantom at 80-140 kVp. Image quality was investigated on image portions in the zones adjacent (A zone) to and distant (D zone) from a bismuth shield. Surface radiation doses were measured using thermoluminescent dosimeters. Streak artefacts were graded on a 3-point-scale. RESULTS:Changes in CTN caused by a bismuth shield resulted in changes in X-ray spectra. CTN changes were more apparent in the A zone than in the D zone, particularly for a low tube voltage. The degrees of CTN changes and image noise were proportional to the thickness of the bismuth shields. A 1-ply bismuth shield reduced surface radiation doses by 7.2%-15.5%. The overall CNRs were slightly degraded, and streak artefacts were acceptable. CONCLUSIONS:Using a bismuth shield could result in significant CTN changes and perceivable artefacts, particularly for a superficial organ close to the shield, and is not recommended for quantification CT examinations or follow-up CT examinations.

SUBMITTER: Liao YL 

PROVIDER: S-EPMC6888723 | biostudies-literature | 2019 Oct

REPOSITORIES: biostudies-literature

altmetric image

Publications

Bismuth shield affecting CT image quality and radiation dose in adjacent and distant zones relative to shielding surface: A phantom study.

Liao Ying-Lan YL   Lai Nan-Ku NK   Tyan Yeu-Sheng YS   Tsai Hui-Yu HY  

Biomedical journal 20191001 5


<h4>Background</h4>To quantify image quality and radiation doses in regions adjacent to and distant from bismuth shields in computed tomography (CT).<h4>Methods</h4>An American College of Radiology accreditation phantom with four solid rods embedded in a water-like background was scanned to verify CT number (CTN) accuracy when using bismuth shields. CTNs, image noise, and contrast-to-noise ratios (CNRs) were determined in the phantom at 80-140 kVp. Image quality was investigated on image portion  ...[more]

Similar Datasets

| S-EPMC9159376 | biostudies-literature
| S-EPMC8471621 | biostudies-literature
| S-EPMC10457930 | biostudies-literature
| S-EPMC4383045 | biostudies-literature
| S-EPMC4666484 | biostudies-literature
| S-EPMC9762351 | biostudies-literature
| PRJEB74081 | ENA
| S-EPMC9386847 | biostudies-literature
| S-EPMC3113580 | biostudies-literature
| S-EPMC8237526 | biostudies-literature