Unknown

Dataset Information

0

Differences in attenuation pattern in myocardial SPECT between CZT and conventional gamma cameras.


ABSTRACT:

Background

In myocardial perfusion imaging (MPI), single-photon emission tomography (SPECT) soft-tissue attenuation by the abdomen, breasts, and lateral chest wall may create artifacts that mimic true perfusion defects. This may cause misdiagnosis of myocardial perfusion. The aim of the present study was to compare the localization, extent, and depth of attenuation artifacts in MPI SPECT for a multi-pinhole cadmium zinc telluride (CZT) camera vs a conventional gamma camera.

Methods

Phantom and patient measurements were performed using a CZT camera (GE NM 530c) and a conventional gamma camera (GE Ventri). All images were attenuation corrected with externally acquired low-dose computed tomography. The localization, extent, and depth of the attenuation artifact were quantified by comparing attenuation-corrected and non-attenuation-corrected images.

Results

Attenuation artifacts were shifted from the inferolateral wall to the lateral wall using the CZT camera compared to a conventional camera in both the patient and the phantom. The extent of the attenuation artifact was significantly larger for the CZT camera compared to the conventional camera (23?±?5% vs 15?±?5%, P?ConclusionsAttenuation artifacts are found in different locations to different extents and depths when using a CZT camera vs a conventional gamma camera for MPI SPECT. This should be taken into consideration when evaluating MPI SPECT studies to avoid misinterpretation of myocardial perfusion distribution.

SUBMITTER: Oddstig J 

PROVIDER: S-EPMC6908561 | biostudies-literature | 2019 Dec

REPOSITORIES: biostudies-literature

altmetric image

Publications

Differences in attenuation pattern in myocardial SPECT between CZT and conventional gamma cameras.

Oddstig Jenny J   Martinsson Elin E   Jögi Jonas J   Engblom Henrik H   Hindorf Cecilia C  

Journal of nuclear cardiology : official publication of the American Society of Nuclear Cardiology 20180523 6


<h4>Background</h4>In myocardial perfusion imaging (MPI), single-photon emission tomography (SPECT) soft-tissue attenuation by the abdomen, breasts, and lateral chest wall may create artifacts that mimic true perfusion defects. This may cause misdiagnosis of myocardial perfusion. The aim of the present study was to compare the localization, extent, and depth of attenuation artifacts in MPI SPECT for a multi-pinhole cadmium zinc telluride (CZT) camera vs a conventional gamma camera.<h4>Methods</h  ...[more]

Similar Datasets

| S-EPMC6581225 | biostudies-literature
| S-EPMC7042031 | biostudies-literature
| S-EPMC5869883 | biostudies-other
| S-EPMC8497048 | biostudies-literature
| S-EPMC3469367 | biostudies-literature
| S-EPMC8212953 | biostudies-literature
| S-EPMC8489737 | biostudies-literature
| S-EPMC7642149 | biostudies-literature
| S-EPMC7343677 | biostudies-literature
| S-EPMC3829891 | biostudies-other