Unknown

Dataset Information

0

Small-field dosimetry of TrueBeamTM flattened and flattening filter-free beams: A multi-institutional analysis.


ABSTRACT: PURPOSE:Detector-dependent interinstitutional variations of the beam data may lead to uncertainties of the delivered dose to patients. Here we evaluated the inter-unit variability of the flattened and flattening filter-free (FFF) beam data of multiple TrueBeam (Varian Medical Systems) linear accelerators focusing on the small-field dosimetry. METHODS:The beam data of 6- and 10-MV photon beams with and without flattening filter measured for modeling of an iPLAN treatment planning system (BrainLAB) were collected from 12 institutions - ten HD120 Multileaf Collimator (MLC) and two Millennium120 MLC. Percent-depth dose (PDD), off-center ratio (OCR), and detector output factors (OFdet ) measured with different detectors were evaluated. To investigate the detector-associated effects, we evaluated the inter-unit variations of the OFdet before and after having applied the output correction factors provided by the International Atomic Energy Agency (IAEA) Technical Reports Series no. 483. RESULTS:PDD measured with a field size of 5 × 5 mm2 showed that the data measured using an ionization chamber had variations exceeding 1% from the median values. The maximum difference from median value was 2.87% for 10 MV photon beam. The maximum variations of the penumbra width for OCR with 10 × 10 mm2 field size were 0.97 mm. The OFdet showed large variations exceeding 15% for a field size of 5 × 5 mm2 . When the output correction factors were applied to the OFdet , the variations were greatly reduced. The relative difference of almost all field output factors were within ± 5% from the median field output factors. CONCLUSION:In this study, the inter-unit variability of small-field dosimetry was evaluated for TrueBeam linear accelerators. The variations were large at a field size of 5 × 5 mm2 , and most occurred in a detector-dependent manner.

SUBMITTER: Akino Y 

PROVIDER: S-EPMC6964782 | biostudies-literature | 2020 Jan

REPOSITORIES: biostudies-literature

altmetric image

Publications

Small-field dosimetry of TrueBeam<sup>TM</sup> flattened and flattening filter-free beams: A multi-institutional analysis.

Akino Yuichi Y   Mizuno Hirokazu H   Isono Masaru M   Tanaka Yoshihiro Y   Masai Norihisa N   Yamamoto Toshijiro T  

Journal of applied clinical medical physics 20191209 1


<h4>Purpose</h4>Detector-dependent interinstitutional variations of the beam data may lead to uncertainties of the delivered dose to patients. Here we evaluated the inter-unit variability of the flattened and flattening filter-free (FFF) beam data of multiple TrueBeam (Varian Medical Systems) linear accelerators focusing on the small-field dosimetry.<h4>Methods</h4>The beam data of 6- and 10-MV photon beams with and without flattening filter measured for modeling of an iPLAN treatment planning s  ...[more]

Similar Datasets

| S-EPMC9861147 | biostudies-literature
| S-EPMC5690198 | biostudies-literature
| S-EPMC5690903 | biostudies-other
| S-EPMC8273795 | biostudies-literature
| S-EPMC5658877 | biostudies-other
| S-EPMC7807632 | biostudies-literature
| S-EPMC5875462 | biostudies-literature
| S-EPMC4266613 | biostudies-literature
| S-EPMC6964765 | biostudies-literature
| S-EPMC5690509 | biostudies-literature