ABSTRACT: Objectives
The purpose of this study was to describe our experience with fluorine-18 fluorodeoxyglucose (FDG) positron emission tomography computed tomography (PET/CT) in diagnosing left ventricular assist device (LVAD) infections and perform a meta-analysis of published studies to determine overall diagnostic accuracy.Background
Device-related infections are a common complication of LVADs and are linked to worse outcomes. Diagnosis of LVAD infections remains challenging. FDG PET/CT has demonstrated good diagnostic accuracy in several other infectious conditions.Methods
This was a single-center, retrospective case series of FDG PET/CT scans in suspected LVAD infection between September 2015 and February 2018. A systematic review of PubMed from database inception through March 2018 was also conducted to identify additional studies.Results
Nineteen FDG PET/CT scans were identified for the retrospective case series. The systematic review identified an additional 3 publications, for a total of 4 studies involving 119 scans assessing diagnostic performance. Axial (n = 36) and centrifugal (n = 83) flow LVADs were represented. Pooled sensitivity was 92% (95% confidence interval [CI]: 82% to 97%) and specificity was 83% (95% CI: 24% to 99%) for FDG PET/CT in diagnosing LVAD infections. Summary receiver-operating characteristic curve analysis demonstrated an AUC of 0.94 (95% CI: 0.91 to 0.95).Conclusions
FDG PET/CT for suspected LVAD infections demonstrates good diagnostic accuracy, with overall high sensitivity but variable specificity.