Lesion Revascularisation Subsequent to Femoropopliteal Spot Stenting Using the Multi-LOC Stent Delivery System.


ABSTRACT: BACKGROUND/AIM:Restenosis and stent fractures are well-characterised treatment failures following femoropopliteal (FP) stent-based interventions. We aimed to determine patterns of restenosis and fracture rates following focal stenting of FP arteries. PATIENTS AND METHODS:This retrospective study investigated angiographic patterns of restenosis and the occurrence of stent fractures following focal FP stenting with the multiple stent delivery system VascuFlex® Multi-LOC. RESULTS:We identified 10 patients with 10 (of 129) FP lesions (mean length 11.7±4.6 cm) and a total of 51 (of 646) Multi-LOC stents that underwent clinically driven target lesion revascularizations (TLR) after 11.5±9.2 months, due to symptomatic recurrence of stenosis. None of the Multi-LOC stents had fractured. No isolated in-stent restenosis (>50%) was observed. Median diameter lumen loss was significantly more pronounced at the inter-stent segments (27.0%) compared to in-stent segments (7.8%, p=0.023). CONCLUSION:No evidence of fracture or susceptibility to stent-related restenosis using Multi-LOC stents was observed. Focal FP stenting may be more effective when combined with strategies known to reduce restenosis in non-stented artery segments.

SUBMITTER: Sigl M 

PROVIDER: S-EPMC6984117 | biostudies-literature | 2020 Jan-Feb

REPOSITORIES: biostudies-literature

altmetric image

Publications

Lesion Revascularisation Subsequent to Femoropopliteal Spot Stenting Using the Multi-LOC Stent Delivery System.

Sigl Martin M   Amendt Klaus K   Waliszewski Matthias M   Rathmann Nils N  

In vivo (Athens, Greece) 20200101 1


<h4>Background/aim</h4>Restenosis and stent fractures are well-characterised treatment failures following femoropopliteal (FP) stent-based interventions. We aimed to determine patterns of restenosis and fracture rates following focal stenting of FP arteries.<h4>Patients and methods</h4>This retrospective study investigated angiographic patterns of restenosis and the occurrence of stent fractures following focal FP stenting with the multiple stent delivery system VascuFlex® Multi-LOC.<h4>Results<  ...[more]

Similar Datasets

| S-EPMC6322090 | biostudies-literature
| S-EPMC7890277 | biostudies-literature
| S-EPMC3932206 | biostudies-other
| S-EPMC9973589 | biostudies-literature
| S-EPMC4743556 | biostudies-literature
| S-EPMC7487302 | biostudies-literature
| S-EPMC8330618 | biostudies-literature
| S-EPMC10478020 | biostudies-literature
| S-EPMC4226912 | biostudies-literature