Synchronized Biventricular Heart Pacing in a Closed-chest Porcine Model based on Wirelessly Powered Leadless Pacemakers.


ABSTRACT: About 30% of patients with impaired cardiac function have ventricular dyssynchrony and seek cardiac resynchronization therapy (CRT). In this study, we demonstrate synchronized biventricular (BiV) pacing in a leadless fashion by implementing miniaturized and wirelessly powered pacemakers. With their flexible form factors, two pacemakers were implanted epicardially on the right and left ventricles of a porcine model and were inductively powered at 13.56?MHz and 40.68?MHz industrial, scientific, and medical (ISM) bands, respectively. The power consumption of these pacemakers is reduced to µW-level by a novel integrated circuit design, which considerably extends the maximum operating distance. Leadless BiV pacing is demonstrated for the first time in both open-chest and closed-chest porcine settings. The clinical outcomes associated with different interventricular delays are verified through electrophysiologic and hemodynamic responses. The closed-chest pacing only requires the external source power of 0.3?W and 0.8?W at 13.56?MHz and 40.68?MHz, respectively, which leads to specific absorption rates (SARs) 2-3 orders of magnitude lower than the safety regulation limit. This work serves as a basis for future wirelessly powered leadless pacemakers that address various cardiac resynchronization challenges.

SUBMITTER: Lyu H 

PROVIDER: S-EPMC7005712 | biostudies-literature | 2020 Feb

REPOSITORIES: biostudies-literature

altmetric image

Publications

Synchronized Biventricular Heart Pacing in a Closed-chest Porcine Model based on Wirelessly Powered Leadless Pacemakers.

Lyu Hongming H   John Mathews M   Burkland David D   Greet Brian B   Post Allison A   Babakhani Aydin A   Razavi Mehdi M  

Scientific reports 20200207 1


About 30% of patients with impaired cardiac function have ventricular dyssynchrony and seek cardiac resynchronization therapy (CRT). In this study, we demonstrate synchronized biventricular (BiV) pacing in a leadless fashion by implementing miniaturized and wirelessly powered pacemakers. With their flexible form factors, two pacemakers were implanted epicardially on the right and left ventricles of a porcine model and were inductively powered at 13.56 MHz and 40.68 MHz industrial, scientific, an  ...[more]

Similar Datasets

| S-EPMC10062368 | biostudies-literature
| S-EPMC10998730 | biostudies-literature
| S-EPMC9986180 | biostudies-literature
| S-EPMC3622721 | biostudies-literature
| S-EPMC6850455 | biostudies-literature
| S-EPMC7319807 | biostudies-literature
| S-EPMC8403316 | biostudies-literature
| S-EPMC4876985 | biostudies-literature
| S-EPMC10285140 | biostudies-literature
| S-EPMC3619489 | biostudies-literature