Large-scale Quality Control of Cardiac Imaging in Population Studies: Application to UK Biobank.


ABSTRACT: In large population studies such as the UK Biobank (UKBB), quality control of the acquired images by visual assessment is unfeasible. In this paper, we apply a recently developed fully-automated quality control pipeline for cardiac MR (CMR) images to the first 19,265 short-axis (SA) cine stacks from the UKBB. We present the results for the three estimated quality metrics (heart coverage, inter-slice motion and image contrast in the cardiac region) as well as their potential associations with factors including acquisition details and subject-related phenotypes. Up to 14.2% of the analysed SA stacks had sub-optimal coverage (i.e. missing basal and/or apical slices), however most of them were limited to the first year of acquisition. Up to 16% of the stacks were affected by noticeable inter-slice motion (i.e. average inter-slice misalignment greater than 3.4?mm). Inter-slice motion was positively correlated with weight and body surface area. Only 2.1% of the stacks had an average end-diastolic cardiac image contrast below 30% of the dynamic range. These findings will be highly valuable for both the scientists involved in UKBB CMR acquisition and for the ones who use the dataset for research purposes.

SUBMITTER: Tarroni G 

PROVIDER: S-EPMC7015892 | biostudies-literature | 2020 Feb

REPOSITORIES: biostudies-literature

altmetric image

Publications

Large-scale Quality Control of Cardiac Imaging in Population Studies: Application to UK Biobank.

Tarroni Giacomo G   Bai Wenjia W   Oktay Ozan O   Schuh Andreas A   Suzuki Hideaki H   Glocker Ben B   Matthews Paul M PM   Rueckert Daniel D  

Scientific reports 20200212 1


In large population studies such as the UK Biobank (UKBB), quality control of the acquired images by visual assessment is unfeasible. In this paper, we apply a recently developed fully-automated quality control pipeline for cardiac MR (CMR) images to the first 19,265 short-axis (SA) cine stacks from the UKBB. We present the results for the three estimated quality metrics (heart coverage, inter-slice motion and image contrast in the cardiac region) as well as their potential associations with fac  ...[more]

Similar Datasets

| S-EPMC2475504 | biostudies-literature
| S-EPMC9454085 | biostudies-literature
| S-EPMC5770339 | biostudies-literature
| S-EPMC7677387 | biostudies-literature
| S-EPMC7641476 | biostudies-literature
| S-EPMC10555669 | biostudies-literature
| S-EPMC6786974 | biostudies-literature
| S-EPMC7060799 | biostudies-literature
| S-EPMC10953622 | biostudies-literature
| S-EPMC7576214 | biostudies-literature