Unknown

Dataset Information

0

Comparison of hemodynamic responses to endotracheal intubation with the GlideScope video laryngoscope and Macintosh laryngoscope in patients undergoing cardiovascular surgery.


ABSTRACT: Background:This study aimed to compare hemodynamic response to endotracheal intubation using GlideScope and Macintosh laryngoscopes in patients who underwent cardiovascular surgery. Methods:A total of 74 patients were enrolled in the study. Patients were randomly assigned into two groups either a GlideScope (n=37) or a Macintosh laryngoscope (n=37). Laryngoscopy time, heart rate, invasive arterial pressure and rate pressure product were compared during induction and until five minutes after laryngoscopy at 11 time points. Results:Seventy patients completed the study. The intubation time was similar in both groups (14.1±4.0 sec vs. 13.2±4.2 sec; p=0.22). Hemodynamic values including heart rate, systolic, diastolic and mean arterial pressure and rate pressure product were similar at all-time points. The laryngoscopy time was also similar between groups. Conclusion:The GlideScope video laryngoscope did not show any advantage on hemodynamic response in patients undergoing cardiovascular surgery when compared to Macintosh laryngoscope.

SUBMITTER: Inangil G 

PROVIDER: S-EPMC7018285 | biostudies-literature | 2018 Jul

REPOSITORIES: biostudies-literature

altmetric image

Publications

Comparison of hemodynamic responses to endotracheal intubation with the GlideScope video laryngoscope and Macintosh laryngoscope in patients undergoing cardiovascular surgery.

İnangil Gökhan G   Cansız Kadir Hakan KH   Gürbüz Fuat F   Bakal Ömer Ö   Gökben Fatma Merih FM   Şen Hüseyin H  

Turk gogus kalp damar cerrahisi dergisi 20180703 3


<h4>Background</h4>This study aimed to compare hemodynamic response to endotracheal intubation using GlideScope and Macintosh laryngoscopes in patients who underwent cardiovascular surgery.<h4>Methods</h4>A total of 74 patients were enrolled in the study. Patients were randomly assigned into two groups either a GlideScope (n=37) or a Macintosh laryngoscope (n=37). Laryngoscopy time, heart rate, invasive arterial pressure and rate pressure product were compared during induction and until five min  ...[more]

Similar Datasets

| S-EPMC9327175 | biostudies-literature