ABSTRACT: Background
This study aimed to compare hemodynamic response to endotracheal intubation using GlideScope and Macintosh laryngoscopes in patients who underwent cardiovascular surgery.Methods
A total of 74 patients were enrolled in the study. Patients were randomly assigned into two groups either a GlideScope (n=37) or a Macintosh laryngoscope (n=37). Laryngoscopy time, heart rate, invasive arterial pressure and rate pressure product were compared during induction and until five minutes after laryngoscopy at 11 time points.Results
Seventy patients completed the study. The intubation time was similar in both groups (14.1±4.0 sec vs. 13.2±4.2 sec; p=0.22). Hemodynamic values including heart rate, systolic, diastolic and mean arterial pressure and rate pressure product were similar at all-time points. The laryngoscopy time was also similar between groups.Conclusion
The GlideScope video laryngoscope did not show any advantage on hemodynamic response in patients undergoing cardiovascular surgery when compared to Macintosh laryngoscope.
SUBMITTER: Inangil G
PROVIDER: S-EPMC7018285 | biostudies-literature | 2018 Jul
REPOSITORIES: biostudies-literature
İnangil Gökhan G Cansız Kadir Hakan KH Gürbüz Fuat F Bakal Ömer Ö Gökben Fatma Merih FM Şen Hüseyin H
Turk gogus kalp damar cerrahisi dergisi 20180703 3
<h4>Background</h4>This study aimed to compare hemodynamic response to endotracheal intubation using GlideScope and Macintosh laryngoscopes in patients who underwent cardiovascular surgery.<h4>Methods</h4>A total of 74 patients were enrolled in the study. Patients were randomly assigned into two groups either a GlideScope (n=37) or a Macintosh laryngoscope (n=37). Laryngoscopy time, heart rate, invasive arterial pressure and rate pressure product were compared during induction and until five min ...[more]