Measuring Blunt Force Head Impacts in Athletes.


ABSTRACT:

Introduction

Although concussion continues to be a major source of acute and chronic injuries, concussion injury mechanisms and risk functions are ill-defined. This lack of definition has hindered efforts to develop standardized concussion monitoring, safety testing, and protective countermeasures. To overcome this knowledge gap, we have developed, tested, and deployed a head impact monitoring mouthguard (IMM) system.

Materials and methods

The IMM system was first calibrated in 731 laboratory tests. Versus reference, Laboratory IMM data fit a linear model, with results close to the ideal linear model of form y = x + 0, R2 = 1. Next, during on-field play involving n = 54 amateur American athletes in football and boxing, there were tens of thousands of events collected by the IMM. A total of 890 true-positive head impacts were confirmed using a combination of signal processing and National Institute of Neurological Disorders and Stroke/National Institutes of Health Common Data Elements methods.

Results

The median and 99th percentile of peak scalar linear acceleration and peak angular acceleration were 20 and 50 g and 1,700 and 4,600 rad/s2, respectively. No athletes were diagnosed with concussion.

Conclusions

While these data are useful for preliminary human tolerance limits, a larger population must be used to quantify real-world dose response as a function of impact magnitude, direction, location, and accumulation. This work is ongoing.

SUBMITTER: Bartsch A 

PROVIDER: S-EPMC7029834 | biostudies-literature | 2020 Jan

REPOSITORIES: biostudies-literature

altmetric image

Publications


<h4>Introduction</h4>Although concussion continues to be a major source of acute and chronic injuries, concussion injury mechanisms and risk functions are ill-defined. This lack of definition has hindered efforts to develop standardized concussion monitoring, safety testing, and protective countermeasures. To overcome this knowledge gap, we have developed, tested, and deployed a head impact monitoring mouthguard (IMM) system.<h4>Materials and methods</h4>The IMM system was first calibrated in 73  ...[more]

Similar Datasets

2024-06-17 | GSE254142 | GEO
2020-04-09 | GSE148325 | GEO
| S-EPMC10996668 | biostudies-literature
| S-EPMC4961456 | biostudies-literature
| S-EPMC10998631 | biostudies-literature
2024-06-17 | GSE254141 | GEO
| S-EPMC3359587 | biostudies-literature
| PRJNA623896 | ENA
| PRJNA1068615 | ENA
2024-06-17 | GSE254138 | GEO