Unknown

Dataset Information

0

Resonant inelastic X-ray scattering using a miniature dispersive Rowland refocusing spectrometer.


ABSTRACT: X-ray absorption spectroscopy (XAS) beamlines worldwide are steadily increasing their emphasis on full photon-in/photon-out spectroscopies, such as resonant inelastic X-ray scattering (RIXS), resonant X-ray emission spectroscopy (RXES) and high energy resolution fluorescence detection XAS (HERFD-XAS). In such cases, each beamline must match the choice of emission spectrometer to the scientific mission of its users. Previous work has recently reported a miniature tender X-ray spectrometer using a dispersive Rowland refocusing (DRR) geometry that functions with high energy resolution even with a large X-ray spot size on the sample [Holden et al. (2017). Rev. Sci. Instrum. 88, 073904]. This instrument has been used in the laboratory in multiple studies of non-resonant X-ray emission spectroscopy using a conventional X-ray tube, though only for preliminary measurements at a low-intensity microfocus synchrotron beamline. This paper reports an extensive study of the performance of a miniature DRR spectrometer at an unfocused wiggler beamline, where the incident monochromatic flux allows for resonant studies which are impossible in the laboratory. The results support the broader use of the present design and also suggest that the DRR method with an unfocused beam could have important applications for materials with low radiation damage thresholds and that would not survive analysis on focused beamlines.

SUBMITTER: Ditter AS 

PROVIDER: S-EPMC7064111 | biostudies-literature | 2020 Mar

REPOSITORIES: biostudies-literature

altmetric image

Publications

Resonant inelastic X-ray scattering using a miniature dispersive Rowland refocusing spectrometer.

Ditter Alexander S AS   Holden William M WM   Cary Samantha K SK   Mocko Veronika V   Latimer Matthew J MJ   Nelson Erik J EJ   Kozimor Stosh A SA   Seidler Gerald T GT  

Journal of synchrotron radiation 20200220 Pt 2


X-ray absorption spectroscopy (XAS) beamlines worldwide are steadily increasing their emphasis on full photon-in/photon-out spectroscopies, such as resonant inelastic X-ray scattering (RIXS), resonant X-ray emission spectroscopy (RXES) and high energy resolution fluorescence detection XAS (HERFD-XAS). In such cases, each beamline must match the choice of emission spectrometer to the scientific mission of its users. Previous work has recently reported a miniature tender X-ray spectrometer using a  ...[more]

Similar Datasets

| S-EPMC6726621 | biostudies-literature
| S-EPMC5792644 | biostudies-literature
| S-EPMC9576770 | biostudies-literature
| S-EPMC6056525 | biostudies-literature
| S-EPMC8935368 | biostudies-literature
| S-EPMC5563841 | biostudies-literature
| S-EPMC6399250 | biostudies-literature
| S-EPMC10352240 | biostudies-literature
| S-EPMC3920600 | biostudies-literature
| S-EPMC4726252 | biostudies-literature