Inter- and intra-reader reproducibility of shear wave elastography measurements for musculoskeletal soft tissue masses.


ABSTRACT:

Objective

To determine inter- and intra-reader reproducibility of shear wave elastography measurements for musculoskeletal soft tissue masses.

Materials and methods

In all, 64 patients with musculoskeletal soft tissue masses were scanned by two readers prior to biopsy; each taking five measurements of shear wave velocity (m/s) and stiffness (kPa). A single lesion per patient was scanned in transverse and cranio-caudal planes. Depth measurements (cm) and volume (cm3) were recorded for each lesion, for each reader. Linear mixed modelling was performed to assess limits of agreement (LOA), inter- and intra-reader repeatability, including analyses for measured depth and volume.

Results

Of the 64 lesions scanned, 24 (38%) were malignant. Bland-Altman plots demonstrated negligible bias with wide LOA for all measurements. Transverse velocity was the most reliable measure-intraclass correlation (95% CI)?=?0.917 (0.886, 1)-though reader 1 measures could be between 38% lower and 57% higher than reader 2 [ratio-scale bias (95% LOA)?=?0.99 (0.64, 1.55)]. Repeatability coefficients indicated most disagreement resulted from poor within-reader reproducibility. LOA between readers calculated from means of five repeated measurements were narrower-transverse velocity ratio-scale bias (95% LOA)?=?1.00 (0.74, 1.35). Depth affected both estimated velocity and repeatability; volume also affected repeatability.

Conclusion

This study found poor repeatability of measurements with wide LOA due mostly to intra-reader variability. Transverse velocity was the most reliable measure; variability may be affected by lesion depth. At least five measurements should be reported with LOA to assist future comparability between shear wave elastography systems in evaluating soft tissue masses.

SUBMITTER: Nicholls J 

PROVIDER: S-EPMC7083807 | biostudies-literature | 2020 May

REPOSITORIES: biostudies-literature

altmetric image

Publications

Inter- and intra-reader reproducibility of shear wave elastography measurements for musculoskeletal soft tissue masses.

Nicholls Jonathan J   Alfuraih Abdulrahman M AM   Hensor Elizabeth M A EMA   Robinson Philip P  

Skeletal radiology 20191212 5


<h4>Objective</h4>To determine inter- and intra-reader reproducibility of shear wave elastography measurements for musculoskeletal soft tissue masses.<h4>Materials and methods</h4>In all, 64 patients with musculoskeletal soft tissue masses were scanned by two readers prior to biopsy; each taking five measurements of shear wave velocity (m/s) and stiffness (kPa). A single lesion per patient was scanned in transverse and cranio-caudal planes. Depth measurements (cm) and volume (cm<sup>3</sup>) wer  ...[more]

Similar Datasets

| S-EPMC6795636 | biostudies-literature
| S-EPMC3654933 | biostudies-literature
| S-EPMC5539095 | biostudies-other
| S-EPMC7066781 | biostudies-literature
| S-EPMC9179262 | biostudies-literature
| S-EPMC3799848 | biostudies-literature
| S-EPMC7103049 | biostudies-literature
| S-EPMC5433696 | biostudies-literature
| S-EPMC5269522 | biostudies-literature
| S-EPMC10073853 | biostudies-literature