ABSTRACT: Purpose
In this paper we compared two different 3D ultrasound (US) modes (3D free-hand mode and 3D wobbler mode) to see which is more suitable to perform the 3D-US/3D-US registration for clinical guidance applications. The typical errors with respect to their impact on the final localization error were evaluated step by step.Methods
Multi-point target and Hand-eye calibration methods were used for 3D US calibration together with a newly designed multi-cone phantom. Pointer based and image based methods were used for 2D US calibration. The calibration target error was computed by using a different multi-cone phantom. An egg-shaped phantom was used as ground truth to compare distortions for both 3D modes along with the measurements of the volume. Finally, we compared 3D ultra