Early stabilization of the uncemented Symax hip stem in a 2-year RSA study.


ABSTRACT: Background and purpose - The uncemented Symax hip stem has shown early proximal ingrowth as result of the BONIT-hydroxyapatite (HA) coating and the distal DOTIZE surface treatment. We evaluated 2-year postoperative radiostereometric analysis (RSA) migration of the Symax hip stem in THA patients. We also investigated the correlation between migration at 4 weeks and clinical outcomes after 2 years.Patients and methods - Patients in a 2-year clinical follow-up single-centre RSA randomized controlled trial were randomized to 2 different cup designs. All 45 patients received a Symax hip stem. RSA migration patterns of the Symax hip stem is presented here as a single cohort. RSA examinations were performed postoperatively, but before weight-bearing, and subsequently after 1, 3, 6, 12, and 24 mon