Tracked Foley catheter for motion compensation during fusion image-guided prostate procedures: a phantom study.


ABSTRACT: BACKGROUND:Uncorrected patient or prostate motion may impair targeting prostate areas during fusion image-guided procedures. We evaluated if a prototype "tracked Foley catheter" (TFC) could maintain fusion image alignment after simulated organ motion. METHODS:A pelvic phantom model underwent magnetic resonance imaging (MRI), and the prostate was segmented. The TFC was placed in the phantom. MRI/ultrasound (US) fusion was performed. Four trials were performed varying motion and TFC presence/absence: (1) TFC/no-motion, (2) TFC/motion, (3) no-TFC/no-motion, and (4) no-TFC/motion. To quantify image alignment, screen captures generated Dice similarity coefficient (DSC) and offset distances (ODs) (maximal US-to-MRI distance between edges on fusion images). Three anatomical targets were identified for placement of a needle under fusion guidance. A computed tomography scan was used to measure system error (SE), i.e., the distance from needle tip to intended target. RESULTS:The TFC presence improved MRI/US alignment by DSC 0.88, 0.88, 0.74, and 0.61 in trials 1, 2, 3, and 4, respectively. Both OD (trial 2 versus trial 4, 4.85 ± 1.60 versus 25.29 ± 6.50?mm, p < 0.001) and SE (trial 2 versus trial 4, 6.35 ± 1.31 versus 32.16 ± 6.50?mm, p < 0.005) were significantly lower when the TFC was present after artificial motion, and significantly smaller OD when static (trial 1 versus trial 3, 4.29 ± 1.24 versus 6.42 ± 2.29?mm, p < 0.001). CONCLUSION:TFC provided better image alignment with or without simulated motion. This may overcome system limitations, allowing for more accurate fusion image alignment during fusion-guided biopsy, ablation, or robotic prostatectomy.

SUBMITTER: Hale GR 

PROVIDER: S-EPMC7163002 | biostudies-literature | 2020 Apr

REPOSITORIES: biostudies-literature

altmetric image

Publications

Tracked Foley catheter for motion compensation during fusion image-guided prostate procedures: a phantom study.

Hale Graham R GR   Pesapane Filippo F   Xu Sheng S   Bakhutashvili Ivane I   Glossop Neil N   Turkbey Baris B   Pinto Peter A PA   Wood Bradford J BJ  

European radiology experimental 20200416 1


<h4>Background</h4>Uncorrected patient or prostate motion may impair targeting prostate areas during fusion image-guided procedures. We evaluated if a prototype "tracked Foley catheter" (TFC) could maintain fusion image alignment after simulated organ motion.<h4>Methods</h4>A pelvic phantom model underwent magnetic resonance imaging (MRI), and the prostate was segmented. The TFC was placed in the phantom. MRI/ultrasound (US) fusion was performed. Four trials were performed varying motion and TFC  ...[more]

Similar Datasets

| S-EPMC7108728 | biostudies-literature
| S-EPMC6941738 | biostudies-literature
| S-EPMC9971291 | biostudies-literature
| S-EPMC6393544 | biostudies-literature
| S-EPMC2771765 | biostudies-literature
| S-EPMC3292926 | biostudies-literature
| S-EPMC4526517 | biostudies-literature
| S-EPMC7063285 | biostudies-literature
| S-EPMC9196943 | biostudies-literature
| S-EPMC9732704 | biostudies-literature