Effect of Tooth Types on the Accuracy of Dental 3D Scanners: An In Vitro Study.
Ontology highlight
ABSTRACT: The purpose of this study was to evaluate the accuracy of dental three-dimensional (3D) scanners according to the types of teeth. A computer-aided design (CAD) reference model (CRM) was obtained by scanning the reference typodont model using a high-precision industrial scanner (Solutionix C500, MEDIT). In addition, a CAD test model (CTM) was obtained using seven types of dental 3D scanners (desktop scanners (E1 and DOF Freedom HD) and intraoral scanners (CS3500, CS3600, Trios2, Trios3, and i500)). The 3D inspection software (Geomagic control X, 3DSystems) was used to segment the CRM according to the types of teeth and to superimpose the CTM based on the segmented teeth. The 3D accuracy of the scanner was then analyzed according to the types of teeth. One-way analysis of variance (ANOVA) was used to compare the differences according to the types of teeth in statistical analysis, and the Tukey HSD test was used for post hoc testing (? = 0.05). Both desktop and intraoral scanners showed significant differences in accuracy according to the types of teeth (P < 0.001), and the accuracy of intraoral scanners tended to get worse from anterior to posterior. Therefore, when scanning a complete arch using an intraoral scanner, the clinician should consider the tendency for the accuracy to decrease from anterior to posterior.
SUBMITTER: Son K
PROVIDER: S-EPMC7178641 | biostudies-literature | 2020 Apr
REPOSITORIES: biostudies-literature
ACCESS DATA