Unknown

Dataset Information

0

Technological quality requirements for stereotactic radiotherapy : Expert review group consensus from the DGMP Working Group for Physics and Technology in Stereotactic Radiotherapy.


ABSTRACT: This review details and discusses the technological quality requirements to ensure the desired quality for stereotactic radiotherapy using photon external beam radiotherapy as defined by the DEGRO Working Group Radiosurgery and Stereotactic Radiotherapy and the DGMP Working Group for Physics and Technology in Stereotactic Radiotherapy. The covered aspects of this review are 1) imaging for target volume definition, 2) patient positioning and target volume localization, 3) motion management, 4) collimation of the irradiation and beam directions, 5) dose calculation, 6) treatment unit accuracy, and 7) dedicated quality assurance measures. For each part, an expert review for current state-of-the-art techniques and their particular technological quality requirement to reach the necessary accuracy for stereotactic radiotherapy divided into intracranial stereotactic radiosurgery in one single fraction (SRS), intracranial fractionated stereotactic radiotherapy (FSRT), and extracranial stereotactic body radiotherapy (SBRT) is presented. All recommendations and suggestions for all mentioned aspects of stereotactic radiotherapy are formulated and related uncertainties and potential sources of error discussed. Additionally, further research and development needs in terms of insufficient data and unsolved problems for stereotactic radiotherapy are identified, which will serve as a basis for the future assignments of the DGMP Working Group for Physics and Technology in Stereotactic Radiotherapy. The review was group peer-reviewed, and consensus was obtained through multiple working group meetings.

SUBMITTER: Schmitt D 

PROVIDER: S-EPMC7182540 | biostudies-literature | 2020 May

REPOSITORIES: biostudies-literature

altmetric image

Publications

Technological quality requirements for stereotactic radiotherapy : Expert review group consensus from the DGMP Working Group for Physics and Technology in Stereotactic Radiotherapy.

Schmitt Daniela D   Blanck Oliver O   Gauer Tobias T   Fix Michael K MK   Brunner Thomas B TB   Fleckenstein Jens J   Loutfi-Krauss Britta B   Manser Peter P   Werner Rene R   Wilhelm Maria-Lisa ML   Baus Wolfgang W WW   Moustakis Christos C  

Strahlentherapie und Onkologie : Organ der Deutschen Rontgengesellschaft ... [et al] 20200324 5


This review details and discusses the technological quality requirements to ensure the desired quality for stereotactic radiotherapy using photon external beam radiotherapy as defined by the DEGRO Working Group Radiosurgery and Stereotactic Radiotherapy and the DGMP Working Group for Physics and Technology in Stereotactic Radiotherapy. The covered aspects of this review are 1) imaging for target volume definition, 2) patient positioning and target volume localization, 3) motion management, 4) co  ...[more]

Similar Datasets

| S-EPMC11343978 | biostudies-literature
| S-EPMC7176344 | biostudies-literature
| S-EPMC7182639 | biostudies-literature
| S-EPMC5834038 | biostudies-literature
| S-EPMC4886133 | biostudies-literature
| S-EPMC6536849 | biostudies-literature
| S-EPMC3656557 | biostudies-literature
| S-EPMC6407034 | biostudies-literature
| S-EPMC4527852 | biostudies-literature
| S-EPMC8128706 | biostudies-literature