New applanation tonometer for myopic patients after laser refractive surgery.
Ontology highlight
ABSTRACT: This study assesses the agreement between intraocular pressure (IOP) measurements taken with the Goldmann applanation tonometer (GAT) and a new experimental applanation tonometer with a convexly shaped apex (CT) after laser myopic refractive surgery. Two different CT radii (CT1 and CT2) were designed with a finite element analyser, and a prospective double masked study on 102 eyes from 102 patients was carried out. A Bland-Altman plot and intra-class correlation coefficient (ICC) were calculated to assess the agreement between GAT measurements and the measurements of both CT1 and CT2 before and after myopic laser assisted in situ keratomileusis (LASIK; n = 73) and photorefractive keratectomy (PRK; n = 29). We evaluated a subset of two subgroups (n = 36 each) for intra and inter-observer (I