Unknown

Dataset Information

0

Evaluation of the optical performance for aspheric intraocular lenses in relation with tilt and decenter errors.


ABSTRACT: PURPOSE:To evaluate and compare the effect of misalignment and tilt on the optical performance of different aspheric intraocular lens (IOL) designs. METHODS:Three aspheric IOLs with a different quantity of spherical aberration (SA) have been designed and the effect of IOL misalignment and tilt on the imaging quality of an eye model has been numerically assessed using a commercial optical design software. The prototypes have been manufactured by lathe turning and tested in vitro using the same optical bench (PMTF, Lambda-X) that complies with International Organization for Standardization standard 11979-2 requirements. Image quality was evaluated from the modulation transfer functions (MTFs), through-focus modulation transfer functions (TF-MTFs), root mean square (RMS) values of defocus, astigmatism and coma, and images of the United States Air Force (USAF) target were taken. A comparison with the optical performance of spherical IOLs has also been performed. RESULTS:Intraocular lens misalignment and tilt increased wavefront aberrations; the effect of misalignment on root mean square (RMS) astigmatism and coma was positively correlated with the spherical aberration of the IOL. Aberration-free IOLs showed the highest MTF for all misalignment values and for IOLs with negative SA correction the MTF decays below 0.43 when they are decentered 0.50 mm. CONCLUSIONS:Aspherical IOLs are more sensitive than spherical IOLs to misalignment or tilt, depending on their SA correction. The optical degradation caused by IOL misalignment had a greater effect on IOL designs with a higher amount of negative spherical aberration. In contrast, the effect of tilt on the optical performance was less sensitive to the IOL design.

SUBMITTER: Perez-Gracia J 

PROVIDER: S-EPMC7197786 | biostudies-literature | 2020

REPOSITORIES: biostudies-literature

altmetric image

Publications

Evaluation of the optical performance for aspheric intraocular lenses in relation with tilt and decenter errors.

Pérez-Gracia Jesús J   Varea Alejandra A   Ares Jorge J   Vallés Juan A JA   Remón Laura L  

PloS one 20200504 5


<h4>Purpose</h4>To evaluate and compare the effect of misalignment and tilt on the optical performance of different aspheric intraocular lens (IOL) designs.<h4>Methods</h4>Three aspheric IOLs with a different quantity of spherical aberration (SA) have been designed and the effect of IOL misalignment and tilt on the imaging quality of an eye model has been numerically assessed using a commercial optical design software. The prototypes have been manufactured by lathe turning and tested in vitro us  ...[more]

Similar Datasets

| S-EPMC7735572 | biostudies-literature
| S-EPMC7137526 | biostudies-literature
| S-EPMC8673660 | biostudies-literature
| S-EPMC5642893 | biostudies-literature
| S-EPMC5448846 | biostudies-other
| S-EPMC5657996 | biostudies-literature
| S-EPMC6516737 | biostudies-literature