Feasibility of Attachable Ring Stimulator for Intraoperative Neuromonitoring during Thyroid Surgery.


ABSTRACT: Objective:Stimulator-attached dissecting instruments are useful for intraoperative nerve monitoring during thyroidectomy. The aim of this study was to evaluate the feasibility of an attachable ring stimulator (ARS) by comparing the electromyography (EMG) amplitudes evoked by an ARS and a conventional stimulator. Methods:Medical records of fourteen patients who underwent thyroidectomy using intraoperative neuromonitoring between June and August 2019 were retrospectively reviewed. The amplitudes of V1, R1, R2, and V2 signals were checked using both the ARS and a conventional stimulator, at the same point. Results:Both stimulators were tested on 20 recurrent laryngeal nerves (RLNs) and 20 vagus nerves (VNs). In all the nerves, the amplitudes of V1, R1, R2, and V2 were greater than 500??V. The mean amplitudes of V1, R1, R2, and V2 checked with the ARS were 1175, 1432, 1598, and 1279??V, respectively. The mean amplitudes of V1, R1, R2, and V2 checked with the conventional stimulator were 1140, 1425, 1557, and 1217??V, respectively. Difference between amplitudes evoked by the two stimulators for V1, R1, R2, and V2 was 77, 110, 102, and 99??V, respectively. There was no statistical difference in the amplitudes between the two groups for V1, R1, R2, and V2. Conclusion:The ARS transferred electric stimulation as effectively as the conventional stimulator. It is an effective tool for repeated stimulation and facilitates continuous feedback regarding the functional integrity of nerves during thyroid surgery.

SUBMITTER: Kim J 

PROVIDER: S-EPMC7204267 | biostudies-literature | 2020

REPOSITORIES: biostudies-literature

altmetric image

Publications

Feasibility of Attachable Ring Stimulator for Intraoperative Neuromonitoring during Thyroid Surgery.

Kim Jongjin J   Moon Hyeon Jong HJ   Chai Young Jun YJ   Lee Jung-Man JM   Hwang Ki-Tae KT   Wu Che-Wei CW   Dionigi Gianlorenzo G   Kim Hoon Yub HY   Park Kyung Sik KS   Kim Sang Wan SW   Yi Ka Hee KH  

International journal of endocrinology 20200130


<h4>Objective</h4>Stimulator-attached dissecting instruments are useful for intraoperative nerve monitoring during thyroidectomy. The aim of this study was to evaluate the feasibility of an attachable ring stimulator (ARS) by comparing the electromyography (EMG) amplitudes evoked by an ARS and a conventional stimulator.<h4>Methods</h4>Medical records of fourteen patients who underwent thyroidectomy using intraoperative neuromonitoring between June and August 2019 were retrospectively reviewed. T  ...[more]

Similar Datasets

| S-EPMC10249491 | biostudies-literature
| S-EPMC11002475 | biostudies-literature
| S-EPMC10493630 | biostudies-literature
| S-EPMC7832964 | biostudies-literature
| S-EPMC9546920 | biostudies-literature
| S-EPMC7435431 | biostudies-literature
| S-EPMC6893217 | biostudies-literature
| S-EPMC7248612 | biostudies-literature
| S-EPMC6590433 | biostudies-literature