Unknown

Dataset Information

0

A Comparative Analysis of Implants Presenting Different Diameters: Extra-Narrow, Narrow and Conventional.


ABSTRACT: This study aimed at performing a comparative analysis of the fracture resistance of implants, evaluating extra-narrow, narrow, and regular implants. Four groups containing 15 implants each were evaluated. Group 1 (G1): single-piece extra-narrow implants; Group 2 (G2): single-piece narrow implants; Group 3 (G3): Morse taper narrow implants with solid abutments; Group 4 (G4): Morse taper conventional implants with solid abutments. The implants were tested using a universal testing machine for their maximum force limit and their maximum bending moment. After obtaining the data, the Shapiro-Wilk, ANOVA, and Tukey (p < 0.05) statistical tests were applied. Samples from all the groups were analyzed by scanning electron microscopy and Groups 3 and 4 were analyzed by profilometry. The means and the standard deviation values for the maximum force limit (N) and the maximum bending moment (Nmm) were respectively: G1:134.29 N (10.27); G2:300.61 N (24.26); G3:360.64 N (23.34); G4:419.10 N (18.87); G1:1612.02 Nmm (100.6); G2:2945 Nmm (237.97); G3:3530.38 Nmm (228.75); G4:4096.7 Nmm (182.73). The groups behaved statistically different from each other, showing that the smallest diameter implants provided less fracture resistance, both in the tensile strength tests and in the maximum bending moment between all groups. Furthermore, single-piece implants, with 2.5 mm and 3.0 mm diameters, deformed in the implant body region area, rather than in the abutment region.

SUBMITTER: Tuzzolo Neto H 

PROVIDER: S-EPMC7215707 | biostudies-literature | 2020 Apr

REPOSITORIES: biostudies-literature

altmetric image

Publications

A Comparative Analysis of Implants Presenting Different Diameters: Extra-Narrow, Narrow and Conventional.

Tuzzolo Neto Henrique H   Tuzita Alessandra Sayuri AS   Gehrke Sérgio Alexandre SA   de Vasconcellos Moura Renata R   Zaffalon Casati Márcio M   Mikail Melo Mesquita Alfredo A  

Materials (Basel, Switzerland) 20200417 8


This study aimed at performing a comparative analysis of the fracture resistance of implants, evaluating extra-narrow, narrow, and regular implants. Four groups containing 15 implants each were evaluated. Group 1 (G1): single-piece extra-narrow implants; Group 2 (G2): single-piece narrow implants; Group 3 (G3): Morse taper narrow implants with solid abutments; Group 4 (G4): Morse taper conventional implants with solid abutments. The implants were tested using a universal testing machine for thei  ...[more]

Similar Datasets

| S-EPMC7408888 | biostudies-literature
| S-EPMC8047002 | biostudies-literature
| S-EPMC7143173 | biostudies-literature
| S-EPMC6086875 | biostudies-literature
| S-EPMC2919048 | biostudies-literature
2019-01-14 | E-MTAB-7383 | biostudies-arrayexpress
| S-EPMC8346526 | biostudies-literature
| S-EPMC7084739 | biostudies-literature
| S-EPMC4340467 | biostudies-literature
| S-EPMC7016390 | biostudies-literature