Improving PCASL at ultra-high field using a VERSE-guided parallel transmission strategy.


ABSTRACT: PURPOSE:To improve the labeling efficiency of pseudo-continuous arterial spin labeling (PCASL) at 7T using parallel transmission (pTx). METHODS:Five healthy subjects were scanned on an 8-channel-transmit 7T human MRI scanner. Time-of-flight (TOF) angiography was acquired to identify regions of interest (ROIs) around the 4 major feeding arteries to the brain, and B1+ and B0 maps were acquired in the labeling plane for tagging pulse design. Complex weights of the labeling pulses for each of the 8 transmit channels were calculated to produce a homogenous radiofrequency (RF) -shimmed labeling across the ROIs. Variable-Rate Selective Excitation (VERSE) pulses were also implemented as a part of the labeling pulse train. Whole-brain perfusion-weighted images were acquired under conditions of RF shimming, VERSE with RF shimming, and standard circularly polarized (CP) mode. The same subjects were scanned on a 3T scanner for comparison. RESULTS:In simulation, VERSE with RF shimming improved the flip-angles across the ROIs in the labeling plane by 90% compared with CP mode. VERSE with RF shimming improved the temporal signal-to-noise ratio by 375% compared with CP mode, but did not outperform a matched 3T sequence with a matched flip-angle. CONCLUSION:We have demonstrated improved PCASL tagging at 7T using VERSE with RF shimming on a commercial head coil under conservative SAR limits at 7T. However, improvements of 7T over 3T may require strategies with less conservative SAR restrictions.

SUBMITTER: Tong Y 

PROVIDER: S-EPMC7216913 | biostudies-literature | 2020 Aug

REPOSITORIES: biostudies-literature

altmetric image

Publications

Improving PCASL at ultra-high field using a VERSE-guided parallel transmission strategy.

Tong Yan Y   Jezzard Peter P   Okell Thomas W TW   Clarke William T WT  

Magnetic resonance in medicine 20200123 2


<h4>Purpose</h4>To improve the labeling efficiency of pseudo-continuous arterial spin labeling (PCASL) at 7T using parallel transmission (pTx).<h4>Methods</h4>Five healthy subjects were scanned on an 8-channel-transmit 7T human MRI scanner. Time-of-flight (TOF) angiography was acquired to identify regions of interest (ROIs) around the 4 major feeding arteries to the brain, and B1+ and B<sub>0</sub> maps were acquired in the labeling plane for tagging pulse design. Complex weights of the labeling  ...[more]

Similar Datasets

| S-EPMC8346258 | biostudies-literature
| S-EPMC7722062 | biostudies-literature
| S-EPMC4891377 | biostudies-literature
| S-EPMC6024401 | biostudies-literature
| S-EPMC4368671 | biostudies-literature
| S-EPMC5548091 | biostudies-literature
| S-EPMC5839578 | biostudies-literature
| S-EPMC5428977 | biostudies-literature
| S-EPMC7745777 | biostudies-literature
| S-EPMC9446284 | biostudies-literature