Dynamic MRI for articulating joint evaluation on 1.5?T and 3.0?T scanners: setup, protocols, and real-time sequences.


ABSTRACT: Dynamic magnetic resonance imaging (MRI) is a non-invasive method that can be used to increase the understanding of the pathomechanics of joints. Various types of real-time gradient echo sequences used for dynamic MRI acquisition of joints include balanced steady-state free precession sequence, radiofrequency-spoiled sequence, and ultra-fast gradient echo sequence. Due to their short repetition time and echo time, these sequences provide high temporal resolution, a good signal-to-noise ratio and spatial resolution, and soft tissue contrast. The prerequisites of the evaluation of joints with dynamic MRI include suitable patient installation and optimal positioning of the joint in the coil to allow joint movement, sometimes with dedicated coil support. There are currently few recommendations in the literature regarding appropriate protocol, sequence standardizations, and diagnostic criteria for the use of real-time dynamic MRI to evaluate joints. This article summarizes the technical parameters of these sequences from various manufacturers on 1.5?T and 3.0?T MRI scanners. We have reviewed pertinent details of the patient and coil positioning for dynamic MRI of various joints. The indications and limitations of dynamic MRI of joints are discussed.

SUBMITTER: Garetier M 

PROVIDER: S-EPMC7237553 | biostudies-literature | 2020 May

REPOSITORIES: biostudies-literature

altmetric image

Publications

Dynamic MRI for articulating joint evaluation on 1.5 T and 3.0 T scanners: setup, protocols, and real-time sequences.

Garetier Marc M   Borotikar Bhushan B   Makki Karim K   Brochard Sylvain S   Rousseau François F   Ben Salem Douraïed D  

Insights into imaging 20200519 1


Dynamic magnetic resonance imaging (MRI) is a non-invasive method that can be used to increase the understanding of the pathomechanics of joints. Various types of real-time gradient echo sequences used for dynamic MRI acquisition of joints include balanced steady-state free precession sequence, radiofrequency-spoiled sequence, and ultra-fast gradient echo sequence. Due to their short repetition time and echo time, these sequences provide high temporal resolution, a good signal-to-noise ratio and  ...[more]

Similar Datasets

| S-EPMC2997510 | biostudies-literature
| S-EPMC3774721 | biostudies-literature
| S-EPMC8134667 | biostudies-literature
| S-EPMC8484226 | biostudies-literature
| S-EPMC4033112 | biostudies-literature
| S-EPMC3382749 | biostudies-literature
| S-EPMC2704948 | biostudies-other
| S-EPMC3843752 | biostudies-other
| S-EPMC8624877 | biostudies-literature
| S-EPMC2704950 | biostudies-other