Indocyanine green matching phantom for fluorescence-guided surgery imaging system characterization and performance assessment.


ABSTRACT: SIGNIFICANCE:Expanded use of fluorescence-guided surgery with devices approved for use with indocyanine green (ICG) has led to a range of commercial systems available. There is a compelling need to be able to independently characterize system performance and allow for cross-system comparisons. AIM:The goal of this work is to expand on previous proposed fluorescence imaging standard designs to develop a long-term stable phantom that spectrally matches ICG characteristics and utilizes 3D printing technology for incorporating tissue-equivalent materials. APPROACH:A batch of test targets was created to assess ICG concentration sensitivity in the 0.3- to 1000-nM range, tissue-equivalent depth sensitivity down to 6 mm, and spatial resolution with a USAF test chart. Comparisons were completed with a range of systems that have significantly different imaging capabilities and applications, including the Li-Cor® Odyssey, Li-Cor® Pearl, PerkinElmer® Solaris, and Stryker® Spy Elite. RESULTS:Imaging of the ICG-matching phantoms with all four commercially available systems showed the ability to benchmark system performance and allow for cross-system comparisons. The fluorescence tests were able to assess differences in the detectable concentrations of ICG with sensitivity differences >10× for preclinical and clinical systems. Furthermore, the tests successfully assessed system differences in the depth-signal decay rate, as well as resolution performance and image artifacts. The manufacturing variations, photostability, and mechanical design of the tests showed promise in providing long-term stable standards for fluorescence imaging. CONCLUSIONS:The presented ICG-matching phantom provides a major step toward standardizing performance characterization and cross-system comparisons for devices approved for use with ICG. The developed hybrid manufacturing platform can incorporate long-term stable fluorescing agents with 3D printed tissue-equivalent material. Further, long-term testing of the phantom and refinements to the manufacturing process are necessary for future implementation as a widely adopted fluorescence imaging standard.

SUBMITTER: Ruiz AJ 

PROVIDER: S-EPMC7240319 | biostudies-literature | 2020 May

REPOSITORIES: biostudies-literature

altmetric image

Publications

Indocyanine green matching phantom for fluorescence-guided surgery imaging system characterization and performance assessment.

Ruiz Alberto J AJ   Wu Mindy M   LaRochelle Ethan P M EPM   Gorpas Dimitris D   Ntziachristos Vasilis V   Pfefer T Joshua TJ   Pogue Brian W BW  

Journal of biomedical optics 20200501 5


<h4>Significance</h4>Expanded use of fluorescence-guided surgery with devices approved for use with indocyanine green (ICG) has led to a range of commercial systems available. There is a compelling need to be able to independently characterize system performance and allow for cross-system comparisons.<h4>Aim</h4>The goal of this work is to expand on previous proposed fluorescence imaging standard designs to develop a long-term stable phantom that spectrally matches ICG characteristics and utiliz  ...[more]

Similar Datasets

| S-EPMC9865296 | biostudies-literature
| S-EPMC4567763 | biostudies-literature
| S-EPMC7311575 | biostudies-literature
| S-EPMC9601265 | biostudies-literature
| S-EPMC9271809 | biostudies-literature
| S-EPMC10219543 | biostudies-literature
| S-EPMC9138305 | biostudies-literature
| S-EPMC10771464 | biostudies-literature
| S-EPMC6160486 | biostudies-other
| S-EPMC4373659 | biostudies-literature