Optimization of a chest computed tomography protocol for detecting pure ground glass opacity nodules: A feasibility study with a computer-assisted detection system and a lung cancer screening phantom.


ABSTRACT: OBJECTIVE:This study aimed to optimize computed tomography (CT) parameters for detecting ground glass opacity nodules (GGNs) using a computer-assisted detection (CAD) system and a lung cancer screening phantom. METHODS:A lung cancer screening phantom containing 15 artificial GGNs (-630 Hounsfield unit [HU], 2-10 mm) in the left lung was examined with a CT scanner. Three tube voltages of 80, 100, and 120 kVp were used in combination with five tube currents of 25, 50, 100, 200, and 400 mA; additionally, three slice thicknesses of 0.625, 1.25, and 2.5 mm and four reconstruction algorithms of adaptive statistical iterative reconstruction (ASIR-V) of 30, 60, and 90% were used. For each protocol, accuracy of the CAD system was evaluated for nine target GGNs of 6, 8, or 10 mm in size. The cut-off size was set to 5 mm to minimize false positives. RESULTS:Among the 180 combinations of tube voltage, tube current, slice thickness, and reconstruction algorithms, combination of 80 kVp, 200 mA, and 1.25-mm slice thickness with an ASIR-V of 90% had the best performance in the detection of GGNs with six true positives and no false positives. Other combinations had fewer than five true positives. In particular, any combinations with a 0.625-mm slice thickness had 0 true positive and at least one false positive result. CONCLUSION:Low-voltage chest CT with a thin slice thickness and a high iterative reconstruction algorithm improve the detection rate of GGNs with a CAD system in a phantom model, and may have potential in lung cancer screening.

SUBMITTER: Kang S 

PROVIDER: S-EPMC7244125 | biostudies-literature | 2020

REPOSITORIES: biostudies-literature

altmetric image

Publications

Optimization of a chest computed tomography protocol for detecting pure ground glass opacity nodules: A feasibility study with a computer-assisted detection system and a lung cancer screening phantom.

Kang Seongmin S   Kim Tae Hoon TH   Shin Jae Min JM   Han Kyunghwa K   Kim Ji Young JY   Min Baeggi B   Park Chul Hwan CH  

PloS one 20200522 5


<h4>Objective</h4>This study aimed to optimize computed tomography (CT) parameters for detecting ground glass opacity nodules (GGNs) using a computer-assisted detection (CAD) system and a lung cancer screening phantom.<h4>Methods</h4>A lung cancer screening phantom containing 15 artificial GGNs (-630 Hounsfield unit [HU], 2-10 mm) in the left lung was examined with a CT scanner. Three tube voltages of 80, 100, and 120 kVp were used in combination with five tube currents of 25, 50, 100, 200, and  ...[more]

Similar Datasets

| S-EPMC7141214 | biostudies-literature
| S-EPMC7336435 | biostudies-literature
| S-EPMC7354160 | biostudies-literature
| S-EPMC10823549 | biostudies-literature
| S-EPMC5717124 | biostudies-literature
| S-EPMC8778636 | biostudies-literature
| S-EPMC10709040 | biostudies-literature
| S-EPMC10996418 | biostudies-literature
| S-EPMC7330407 | biostudies-literature
| S-EPMC7305264 | biostudies-literature