Diagnostic accuracy of point-of-care devices for detection of anemia in community settings in India.


ABSTRACT:

Background

Accurate diagnosis of anemia by community workers using a point-of-care device is a challenge. The objective of the study was to establish the diagnostic accuracy of point-of-care devices for detecting anemia in community settings.

Methods

It was diagnostic accuracy study with cross-sectional design on adult patients attending the outpatient department of rural/ urban health centres of Medical colleges from India. The index tests were HemoCue, TrueHb, Massimo's device and spectroscopic device, compared against autoanalyzer (gold standard). Accuracy was expressed by sensitivity, specificity, likelihood ratios, predictive values, area under the curve (AUC) and levels of agreement. For the diagnostic accuracy component, 1407 participants were recruited with a minimum of 600 for each device. An additional 200 participants were considered to elucidate the performance of devices in different weather conditions.

Results

HemoCue and TrueHb performed better than Massimo and spectroscopic devices. Detection of anemia by technicians was similar between TrueHb and HemoCue (AUC 0.92 v/s 0.90, p >?0.05). Community workers performed better with Hemocue for detecting anemia compared to TrueHb (AUC 0.92 v/s 0.90, p ConclusionFor detection of anemia, True Hb and HemoCue were comparable. For severe anemia, True Hb seemed to be a better and feasible point-of-care device for detecting anemia in the community settings.

SUBMITTER: Neogi SB 

PROVIDER: S-EPMC7249358 | biostudies-literature | 2020 May

REPOSITORIES: biostudies-literature

altmetric image

Publications

Diagnostic accuracy of point-of-care devices for detection of anemia in community settings in India.

Neogi Sutapa Bandyopadhyay SB   Sharma Jyoti J   Pandey Shivam S   Zaidi Nausheen N   Bhattacharya Maitreyee M   Kar Rakhee R   Kar Sitanshu Sekhar SS   Purohit Abhishek A   Bandyopadhyay Sanjib S   Saxena Renu R  

BMC health services research 20200526 1


<h4>Background</h4>Accurate diagnosis of anemia by community workers using a point-of-care device is a challenge. The objective of the study was to establish the diagnostic accuracy of point-of-care devices for detecting anemia in community settings.<h4>Methods</h4>It was diagnostic accuracy study with cross-sectional design on adult patients attending the outpatient department of rural/ urban health centres of Medical colleges from India. The index tests were HemoCue, TrueHb, Massimo's device a  ...[more]

Similar Datasets

| S-EPMC6190021 | biostudies-literature
| S-EPMC10033918 | biostudies-literature
| S-EPMC9105343 | biostudies-literature
| S-EPMC7752099 | biostudies-literature
| S-EPMC9312898 | biostudies-literature
| S-EPMC7833111 | biostudies-literature
| S-EPMC9207186 | biostudies-literature
| S-EPMC8104425 | biostudies-literature
| S-EPMC5636681 | biostudies-literature
| S-EPMC8808351 | biostudies-literature