Taper fretting corrosion with Stryker Anato stem after hip replacement.


ABSTRACT: We report a case of head-neck taper fretting corrosion in a patient who had a total hip replacement with a noncemented Stryker Anato femoral stem and a V40 metal head with a Stryker Tritanium hemispherical socket with a highly cross-linked polyethylene liner (metal on polyethylene) (Stryker, Mahwah, NJ, USA). A 57-year-old man presented with early-onset hip pain after right total hip arthroplasty. Workup was negative for infection. Metal artifact reduction sequence MRI revealed an encapsulated fluid mass. Metal ion cobalt level was elevated at 6 ppb. The patient underwent right revision total hip arthroplasty with excellent results at 1-year follow-up.

SUBMITTER: Freccero DM 

PROVIDER: S-EPMC7303489 | biostudies-literature | 2020 Jun

REPOSITORIES: biostudies-literature

altmetric image

Publications

Taper fretting corrosion with Stryker Anato stem after hip replacement.

Freccero David M DM   McAlpine Kenneth J KJ   Smith Eric L EL  

Arthroplasty today 20200331 2


We report a case of head-neck taper fretting corrosion in a patient who had a total hip replacement with a noncemented Stryker Anato femoral stem and a V40 metal head with a Stryker Tritanium hemispherical socket with a highly cross-linked polyethylene liner (metal on polyethylene) (Stryker, Mahwah, NJ, USA). A 57-year-old man presented with early-onset hip pain after right total hip arthroplasty. Workup was negative for infection. Metal artifact reduction sequence MRI revealed an encapsulated f  ...[more]

Similar Datasets

| S-EPMC5362300 | biostudies-literature
| S-EPMC6588719 | biostudies-literature
| S-EPMC5712029 | biostudies-literature
| S-EPMC7177599 | biostudies-literature
| S-EPMC9661431 | biostudies-literature
| S-EPMC4958113 | biostudies-literature
| S-EPMC6868773 | biostudies-literature
| S-EPMC4762799 | biostudies-other
| S-EPMC5264576 | biostudies-literature
| S-EPMC5551794 | biostudies-other