To measure T1 of short T2 species using an inversion recovery prepared three-dimensional ultrashort echo time (3D IR-UTE) method: A phantom study.


ABSTRACT:

Purpose

To demonstrate the feasibility of a new method for measuring T1 of short T2 species based on an adiabatic inversion recovery-prepared three-dimensional ultrashort echo time Cones (3D IR-UTE-Cones) sequence.

Methods

T1 values for short T2 species were quantified using 3D IR-UTE-Cones data acquired with different repetition times (TRs) and inversion times (TIs). An inversion efficiency factor Q was introduced into the fitting model to accurately calculate T1 values for short T2 species. Experiments were performed on twelve MnCl2 aqueous solution phantoms with a wide range of T1 values and T2* values on a 3 T clinical MR system to verify the efficacy of the proposed method. For comparison, a variable flip angle UTE (VFA-UTE) sequence, a variable TR UTE (VTR-UTE) sequence, and a conventional 2D IR fast spin echo (IR-FSE) sequence were also used to quantify T1 values of those phantoms. T1 values were compared between all performed sequences.

Results

The proposed 3D IR-UTE-Cones method provided higher contrast images of short T2 phantoms and measured much shorter T1 values than the VFA-UTE, VTR-UTE and 2D IR-FSE methods. T1 values as short as 2.95 ms could be measured by the 3D IR-UTE-Cones sequence. The 3D IR-UTE-Cones methods with different TRs were applied to different ranges of T1 measurement, and the scan time was significantly decreased by using 5 TIs along the recovery curves to perform fitting with comparable accuracy.

Conclusion

The 3D IR-UTE-Cones sequence could accurately measure short T1 values while providing high contrast images of short T2 species.

SUBMITTER: Wei Z 

PROVIDER: S-EPMC7307614 | biostudies-literature | 2020 May

REPOSITORIES: biostudies-literature

altmetric image

Publications

To measure T<sub>1</sub> of short T<sub>2</sub> species using an inversion recovery prepared three-dimensional ultrashort echo time (3D IR-UTE) method: A phantom study.

Wei Zhao Z   Ma Ya-Jun YJ   Jang Hyungseok H   Yang Wenhui W   Du Jiang J  

Journal of magnetic resonance (San Diego, Calif. : 1997) 20200413


<h4>Purpose</h4>To demonstrate the feasibility of a new method for measuring T<sub>1</sub> of short T<sub>2</sub> species based on an adiabatic inversion recovery-prepared three-dimensional ultrashort echo time Cones (3D IR-UTE-Cones) sequence.<h4>Methods</h4>T<sub>1</sub> values for short T<sub>2</sub> species were quantified using 3D IR-UTE-Cones data acquired with different repetition times (TRs) and inversion times (TIs). An inversion efficiency factor Q was introduced into the fitting model  ...[more]

Similar Datasets

| S-EPMC8944187 | biostudies-literature
| S-EPMC4263352 | biostudies-other
| S-EPMC5821562 | biostudies-literature
| S-EPMC6996715 | biostudies-literature
| S-EPMC10973042 | biostudies-literature
| S-EPMC7895481 | biostudies-literature
| S-EPMC11305063 | biostudies-literature
| S-EPMC7643815 | biostudies-literature
| S-EPMC6949380 | biostudies-literature
| S-EPMC7952008 | biostudies-literature