Unknown

Dataset Information

Usefulness of point-of-care multiplex PCR to rapidly identify pathogens responsible for ventilator-associated pneumonia and their resistance to antibiotics: an observational study.


ABSTRACT:

Background

The use of multiplex PCR to shorten time to identification of pathogens and their resistance mechanisms for patients with ventilator-associated pneumonia (VAP) is attractive, but poorly studied. The multiplex PCR-based Unyvero pneumonia cartridge assay can directly identify 20 bacteria and one fungus, amongst the most frequently causing VAP, and 19 of their resistance markers in clinical specimens (bronchoalveolar lavage or tracheal aspirate), with a turnaround time of 4-5 h. We performed this study to evaluate the concordance between the multiplex PCR-based Unyvero pneumonia cartridge assay and conventional microbiological techniques to identify pathogens and their resistance mechanisms in patients with VAP.

Methods

All patients suspected of having VAP (January 2

SUBMITTER: Luyt CE 

PROVIDER: S-EPMC7316635 | biostudies-literature | 2020 Jun