Evaluation of low cryptococcal antigen titer as determined by the lateral flow assay in serum and cerebrospinal fluid among HIV-negative patients: a retrospective diagnostic accuracy study.


ABSTRACT: Cryptococcosis is one of the most common opportunistic infections in both immunocompetent and immunocompromised hosts. Although the cryptococcal antigen (CrAg) lateral flow assay (LFA) has been widely used in clinical settings due to its high sensitivity and specificity, the diagnostic value of a low CrAg LFA titers remains unclear. In this study, we performed a retrospective analysis of 149 HIV-negative patients with low CrAg LFA titers (?1:10) in a Chinese tertiary hospital from January 2013 to December 2017, to evaluate the diagnostic value of low CrAg LFA titers in serum and cerebrospinal fluid (CSF) at different thresholds. Sensitivity and specificity of low CrAg LFA titers in patients with definitive diagnoses of cryptococcosis were 39.6% (95% CI, 29.7-50.1%) and 100% (95% CI, 69.2-100%), respectively, at a threshold of 1:10 in serum. A sensitivity of 72.9% (95% CI, 62.9-81.5%) and a decreased specificity of 70.0% (95% CI, 34.8-93.3%) were observed at a threshold of 1:5 in serum. No false-positive cases were identified in patients with low CrAg titers in CSF and all positive predictive values (PPVs) were 100%. Among the cases with low serum CrAg titers, lumbar puncture was performed in 97 patients and positive CSF CrAg titers were reported in 6 patients. In conclusion, the results of this study imply that low CrAg LFA titer, either in serum or CSF, is crucial for early diagnosis of cryptococcosis in HIV-negative patients, and lumbar puncture is recommended to be performed routinely for CSF testing when a positive low serum titer is reported. Cryptococcal meningitis should be considered seriously when the CSF CrAg titer is positive.

SUBMITTER: Wang X 

PROVIDER: S-EPMC7325107 | biostudies-literature | 2020

REPOSITORIES: biostudies-literature

altmetric image

Publications

Evaluation of low cryptococcal antigen titer as determined by the lateral flow assay in serum and cerebrospinal fluid among HIV-negative patients: a retrospective diagnostic accuracy study.

Wang Xuan X   Cheng Jia-Hui JH   Zhou Ling-Hong LH   Zhu Jun-Hao JH   Wang Rui-Ying RY   Zhao Hua-Zhen HZ   Jiang Ying-Kui YK   Huang Li-Ping LP   Yip Ching-Wan CW   Que Chun-Xing CX   Zhu Min M   Zhu Li-Ping LP  

IMA fungus 20200310


Cryptococcosis is one of the most common opportunistic infections in both immunocompetent and immunocompromised hosts. Although the cryptococcal antigen (CrAg) lateral flow assay (LFA) has been widely used in clinical settings due to its high sensitivity and specificity, the diagnostic value of a low CrAg LFA titers remains unclear. In this study, we performed a retrospective analysis of 149 HIV-negative patients with low CrAg LFA titers (≤1:10) in a Chinese tertiary hospital from January 2013 t  ...[more]