ABSTRACT: Background
In anterior cruciate ligament (ACL) reconstruction, minimizing the graft-tunnel motion (GTM) will promote graft-to-bone healing and avoid graft loosening or tearing as well as potential bone tunnel enlargement. A nearly isometric state of the graft can be achieved by placing the tunnel properly to theoretically gain better graft-to-bone healing. However, little clinical evidence is available to quantify the relation between GTM and tunnel position.Purpose
To find the proper zones for the femoral and tibial tunnel apertures that minimize the GTM, referred to as the "nearly isometric zone," through use of intraoperative GTM measurement and 3-dimensional computed tomography (3D-CT).Study design
Cross-sectional study; Level of evidence, 3.Methods
A total of 100 patients were enrolled in this study. Nearly isometric ACL reconstruction was performed, and an intra-articular GTM measuring device was designed to measure and record the amplitude of GTM while the knee was flexed from 0° to 120°. Postoperatively, the patients underwent multislice CT, and the images were used to create 3D-CT models. After tibial aperture examination, 5 patients were excluded due to the divergence of tibial aperture, and therefore 95 patients remained in the study. Patients were divided into 2 groups according to whether the lateral intercondylar ridge was absent or present. The Bernard-Hertel grid coordinates (h, t) of the femoral tunnel were then quantified.Results
The maximal GTM (mGTM) was a mean ± SD of 1.06 ± 0.66 mm (range, 0.0-3.0 mm). The mGTM in patients with a lateral intercondylar ridge was significantly lower than that in patients without a lateral intercondylar ridge (0.81 ± 0.39 vs 1.59 ± 0.73 mm, respectively; P < .0001). The average h and t were 0.227 ± 0.079 and 0.429 ± 0.770, respectively. Notably, in 1 patient, the mGTM was 0 mm whereas the coordinates (h, t) of the femoral tunnel were 0.250 and 0.255. The overall GTM slowly increased before 90° but increased significantly after the knee was bent 105° (P = .010). Correlation analysis showed that the t coordiinate had significant correlation with mGTM (R = 0.581; P < .001). A gradient pattern was created to show the nearly isometric blue zone (mGTM <0.5 mm), which was found to overlap with the IDEAL (isometric, direct insertion, eccentric, anatomic, low tension-flexion pattern) position.Conclusion
A method of measuring intraoperative GTM and quantifying femoral tunnel position on postoperative 3D-CT was successfully developed. The presence of a lateral condylar ridge can significantly reduce mGTM. A nearly isometric zone was described that was consistent with the IDEAL concept.
SUBMITTER: Wan F
PROVIDER: S-EPMC7333503 | biostudies-literature | 2019 Dec
REPOSITORIES: biostudies-literature
Orthopaedic journal of sports medicine 20191231 12
<h4>Background</h4>In anterior cruciate ligament (ACL) reconstruction, minimizing the graft-tunnel motion (GTM) will promote graft-to-bone healing and avoid graft loosening or tearing as well as potential bone tunnel enlargement. A nearly isometric state of the graft can be achieved by placing the tunnel properly to theoretically gain better graft-to-bone healing. However, little clinical evidence is available to quantify the relation between GTM and tunnel position.<h4>Purpose</h4>To find the p ...[more]