Quantitative Method for Comparative Assessment of Particle Removal Efficiency of Fabric Masks as Alternatives to Standard Surgical Masks for PPE.


ABSTRACT: In response to the COVID-19 pandemic, cloth masks are being used to control the spread of virus, but the efficacy of these loose-fitting masks is not well known. Here, tools and methods typically used to assess tight-fitting respirators were modified to quantify the efficacy of community-produced and commercially produced fabric masks as personal protective equipment. Two particle counters concurrently sample ambient air and air inside the masks; mask performance is evaluated by mean particle removal efficiency and statistical variability when worn as designed and with a nylon overlayer, to independently assess fit and material. Worn as designed, both commercial surgical masks and cloth masks had widely varying effectiveness (53%-75% and 28%-91% particle removal efficiency, respectively). Most surgical-style masks improved with the nylon overlayer, indicating poor fit. This rapid testing method uses widely available hardware, requires only a few calculations from collected data, and provides both a holistic and aspect-wise evaluation of mask performance.

SUBMITTER: Mueller AV 

PROVIDER: S-EPMC7346791 | biostudies-literature | 2020 Sep

REPOSITORIES: biostudies-literature

altmetric image

Publications

Quantitative Method for Comparative Assessment of Particle Removal Efficiency of Fabric Masks as Alternatives to Standard Surgical Masks for PPE.

Mueller Amy V AV   Eden Matthew J MJ   Oakes Jessica M JM   Bellini Chiara C   Fernandez Loretta A LA  

Matter 20200709 3


In response to the COVID-19 pandemic, cloth masks are being used to control the spread of virus, but the efficacy of these loose-fitting masks is not well known. Here, tools and methods typically used to assess tight-fitting respirators were modified to quantify the efficacy of community-produced and commercially produced fabric masks as personal protective equipment. Two particle counters concurrently sample ambient air and air inside the masks; mask performance is evaluated by mean particle re  ...[more]

Similar Datasets

| S-EPMC7939758 | biostudies-literature
| S-EPMC7956512 | biostudies-literature
| S-EPMC8187651 | biostudies-literature
| S-EPMC7556527 | biostudies-literature
| S-EPMC6879156 | biostudies-literature
| S-EPMC3591312 | biostudies-literature
| S-EPMC7534667 | biostudies-literature
| S-EPMC8301741 | biostudies-literature
| S-EPMC9303050 | biostudies-literature