Unknown

Dataset Information

0

Incubator-based Sound Attenuation: Active Noise Control In A Simulated Clinical Environment.


ABSTRACT: OBJECTIVE:Noise in the neonatal intensive care unit can be detrimental to the health of the hospitalized infant. Means of reducing that noise include staff training, warning lights, and ear coverings, all of which have had limited success. Single family rooms, while an improvement, also expose the hospitalized infant to the same device alarms and mechanical noises found in open bay units. METHODS:We evaluated a non-contact incubator-based active noise control device (Neoasis™, Invictus Medical, San Antonio, Texas) in a simulated neonatal intensive care unit (NICU) setting to determine whether it could effectively reduce the noise exposure of infants within an incubator. In the NICU simulation center, we generated a series of clinically appropriate sound sequences with bedside medical devices such as a patient monitor and fluid infusion devices, hospital air handling systems, and device mechanical sounds. A microphone-equipped infant mannequin was oriented within an incubator. Measurements were made with the microphones with the Neoasis™ deactivated and activated. RESULTS:The active noise control device decreased sound pressure levels for certain alarm sounds by as much as 14.4 dB (a 5.2-fold reduction in sound pressure) at the alarm tone's primary frequency. Frequencies below the 2 kHz octave band were more effectively attenuated than frequencies at or above the 2 kHz octave band. Background noise levels below 40 dBA were essentially not impacted by the active noise control device. CONCLUSIONS:The active noise control device further reduces noise inside infant incubators. Device safety and potential health benefits of the quieter environment should be verified in a clinical setting.

SUBMITTER: Hutchinson G 

PROVIDER: S-EPMC7363066 | biostudies-literature | 2020

REPOSITORIES: biostudies-literature

altmetric image

Publications

Incubator-based Sound Attenuation: Active Noise Control In A Simulated Clinical Environment.

Hutchinson George G   Du Lilin L   Ahmad Kaashif K  

PloS one 20200715 7


<h4>Objective</h4>Noise in the neonatal intensive care unit can be detrimental to the health of the hospitalized infant. Means of reducing that noise include staff training, warning lights, and ear coverings, all of which have had limited success. Single family rooms, while an improvement, also expose the hospitalized infant to the same device alarms and mechanical noises found in open bay units.<h4>Methods</h4>We evaluated a non-contact incubator-based active noise control device (Neoasis™, Inv  ...[more]

Similar Datasets

| S-EPMC8927378 | biostudies-literature
| S-EPMC7695846 | biostudies-literature
| S-EPMC5876390 | biostudies-literature
| S-EPMC8795453 | biostudies-literature
| S-EPMC7021483 | biostudies-literature
| S-EPMC7264369 | biostudies-literature
| S-EPMC8172042 | biostudies-literature
| S-EPMC7347542 | biostudies-literature
| S-EPMC2861702 | biostudies-literature
| S-EPMC3825667 | biostudies-literature