Development and validation of a machine learning, smartphone-based tonometer.


ABSTRACT:

Background/aims

To compare intraocular pressure (IOP) measurements using a prototype smartphone tonometer with other tonometers used in clinical practice.

Methods

Patients from an academic glaucoma practice were recruited. The smartphone tonometer uses fixed force applanation and in conjunction with a machine-learning computer algorithm is able to calculate the IOP. IOP was also measured using Goldmann applanation tonometry (GAT) in all subjects. A subset of patients were also measured using ICare, pneumotonometry (upright and supine positions) and Tono-Pen (upright and supine positions) and the results were compared.

Results

92 eyes of 81 subjects were successfully measured. The mean difference (in mm Hg) for IOP measurements of the smartphone tonometer versus other