Influence of Orthodontic Anchor Screw Anchorage Method on the Stability of Artificial Bone: An In Vitro Study.


ABSTRACT: This study aims to compare the torque values for various lengths of the titanium-based orthodontic anchor screw (OAS), different anchorage methods and varying artificial bone densities after predrilling. Furthermore, the effects of these parameters on bone stability are evaluated. A total of 144 OASs were prepared with a diameter of 1.6 mm and heights of 6, 8 and 10 mm. Artificial bones were selected according to their density, corresponding to Grades 50, 40 and 30. Torque values for the automatic device and manual anchorage methods exhibited a statistically significant difference for the same-sized OAS, according to the bone density of the artificial bones (p < 0.05). However, when insertion torque was at the maximum rotations, there was no significant difference in the torque values for the Grade 30 artificial bone (p > 0.05). When the torque values of both anchorage methods were statistically compared with the mean difference for each group, the results of the manual anchorage method were significantly higher than those of the automatic device anchorage method (p < 0.05). A statistically significant difference was observed in the bone stability resulting from different OAS anchorage methods and artificial bone lengths. These findings suggest that the automatic anchorage method should be used when fixing the OAS.

SUBMITTER: Kang SY 

PROVIDER: S-EPMC7411596 | biostudies-literature | 2020 Jul

REPOSITORIES: biostudies-literature

altmetric image

Publications

Influence of Orthodontic Anchor Screw Anchorage Method on the Stability of Artificial Bone: An In Vitro Study.

Kang Seen-Young SY   Yu Ji-Min JM   Kim Hyoung-Sik HS   Lee Jun-Seok JS   Yeon Chan-Mi CM   Park Ki-Sook KS   Choi Sung-Hwan SH   Lee Seung-Youl SY  

Materials (Basel, Switzerland) 20200718 14


This study aims to compare the torque values for various lengths of the titanium-based orthodontic anchor screw (OAS), different anchorage methods and varying artificial bone densities after predrilling. Furthermore, the effects of these parameters on bone stability are evaluated. A total of 144 OASs were prepared with a diameter of 1.6 mm and heights of 6, 8 and 10 mm. Artificial bones were selected according to their density, corresponding to Grades 50, 40 and 30. Torque values for the automat  ...[more]

Similar Datasets

| S-EPMC5512929 | biostudies-other
| S-EPMC7125198 | biostudies-literature
| S-EPMC4892476 | biostudies-literature
| S-EPMC8926351 | biostudies-literature
| S-EPMC6921117 | biostudies-literature
| S-EPMC8596623 | biostudies-literature
| S-EPMC6968847 | biostudies-literature
| S-EPMC6464832 | biostudies-literature
| S-EPMC7539479 | biostudies-literature
2017-03-12 | GSE95467 | GEO