Unknown

Dataset Information

0

Depth-Dependent Corneal Biomechanical Properties in Normal and Keratoconic Subjects by Optical Coherence Elastography.


ABSTRACT: Purpose:Compare depth-resolved biomechanical properties in normal and keratoconic corneas in live human subjects using optical coherence elastography (OCE). Methods:In a prospective series of normal and keratoconus (KC) eyes, a corneal perturbation was applied by a custom swept-source OCE system using a transparent flat lens coupled to force transducers. Cross-correlation was applied to track frame-by-frame OCT speckle displacement. Regional displacements for the anterior and posterior stroma were plotted in force versus displacement (k) graphs. A spatial biomechanical property ratio (ka/kp ) was defined by dividing the maximum total displacement by the maximum force for the anterior (ka ) and posterior cornea (kp) and was compared between normal and KC groups with the Mann-Whitney U test. Area under the receiver operating characteristics curve (AUROC) for differentiating normal and KC eyes was calculated for ka/kp , kmax, and thinnest point of corneal thickness (TPCT). Results:Thirty-six eyes were analyzed (21 eyes of 12 normal subjects and 15 KC eyes of 12 subjects). The ka/kp for the normal group was 1.135 ± 0.07 (mean ± standard deviation) and 1.02 ± 0.08 for the KC group (P < 0.001), indicating a relative deficit in anterior stromal stiffness in KC eyes. AUROC was 0.91 for ka /kp , 0.95 for kmax, and 1 for TPCT. Conclusions:Significant differences in depth-dependent corneal biomechanical properties were observed between normal and KC subjects. Translational Relevance:OCE was applied for the first time to human KC subjects and revealed alterations in the normal anterior-to-posterior stromal stiffness gradient, a novel and clinically accessible disease biomarker.

SUBMITTER: De Stefano VS 

PROVIDER: S-EPMC7414661 | biostudies-literature | 2020 Jun

REPOSITORIES: biostudies-literature

altmetric image

Publications

Depth-Dependent Corneal Biomechanical Properties in Normal and Keratoconic Subjects by Optical Coherence Elastography.

De Stefano Vinicius S VS   Ford Matthew R MR   Seven Ibrahim I   Dupps William J WJ  

Translational vision science & technology 20200603 7


<h4>Purpose</h4>Compare depth-resolved biomechanical properties in normal and keratoconic corneas in live human subjects using optical coherence elastography (OCE).<h4>Methods</h4>In a prospective series of normal and keratoconus (KC) eyes, a corneal perturbation was applied by a custom swept-source OCE system using a transparent flat lens coupled to force transducers. Cross-correlation was applied to track frame-by-frame OCT speckle displacement. Regional displacements for the anterior and post  ...[more]

Similar Datasets

| S-EPMC8322708 | biostudies-literature
| S-EPMC7415307 | biostudies-literature
| S-EPMC6252009 | biostudies-other
| S-EPMC3514625 | biostudies-literature
| S-EPMC5316890 | biostudies-other
| S-EPMC6310362 | biostudies-literature
| S-EPMC6097919 | biostudies-literature
| S-EPMC7004960 | biostudies-literature
| S-EPMC5018684 | biostudies-literature
| S-EPMC7154028 | biostudies-literature