Ontology highlight
ABSTRACT:
SUBMITTER: Murthy V
PROVIDER: S-EPMC7447708 | biostudies-literature | 2020 Aug
REPOSITORIES: biostudies-literature
Murthy Vishnu V Smith Raven L RL Tao Dora H DH Lawhn-Heath Courtney A CA Korenchan Dave E DE Larson Peder E Z PEZ Flavell Robert R RR Hope Thomas A TA
EJNMMI physics 20200826 1
<h4>Background</h4>In this study, we investigate the impact of increased PET acquisition time per bed position on lesion detectability, standard uptake value, and image noise in <sup>68</sup>Ga-PSMA-11 PET/MRI scans.<h4>Methods</h4>Scans of twenty patients were analyzed in this study. Patients were injected with <sup>68</sup>Ga-PSMA-11 (mean, 5.50 ± 1.49 mCi) and imaged on a 3.0 T time-of-flight PET/MRI. PET images were retrospectively reconstructed using 0.5, 1, 2, 4, 7, and 10 min of PET data. ...[more]