Improved Testing and Design of Intubation Boxes During the COVID-19 Pandemic.


ABSTRACT:

Study objective

Throughout the coronavirus disease 2019 pandemic, many emergency departments have been using passive protective enclosures ("intubation boxes") during intubation. The effectiveness of these enclosures remains uncertain. We sought to quantify their ability to contain aerosols using industry standard test protocols.

Methods

We tested a commercially available passive protective enclosure representing the most common design and compared this with a modified enclosure that incorporated a vacuum system for active air filtration during simulated intubations and negative-pressure isolation. We evaluated the enclosures by using the same 3 tests air filtration experts use to certify class I biosafety cabinets: visual smoke pattern analysis using neutrally buoyant smoke, aerosol leak testing using a test aerosol that mimics the size of virus-containing particulates, and air velocity measurements.

Results

Qualitative evaluation revealed smoke escaping from all passive enclosure openings. Aerosol leak testing demonstrated elevated particle concentrations outside the enclosure during simulated intubations. In contrast, vacuum-filter-equipped enclosures fully contained the visible smoke and test aerosol to standards consistent with class I biosafety cabinet certification.

Conclusion

Passive enclosures for intubation failed to contain aerosols, but the addition of a vacuum and active air filtration reduced aerosol spread during simulated intubation and patient isolation.

SUBMITTER: Turer DM 

PROVIDER: S-EPMC7470714 | biostudies-literature | 2021 Jan

REPOSITORIES: biostudies-literature

altmetric image

Publications

Improved Testing and Design of Intubation Boxes During the COVID-19 Pandemic.

Turer David M DM   Good Cameron H CH   Schilling Benjamin K BK   Turer Robert W RW   Karlowsky Nicholas R NR   Dvoracek Lucas A LA   Ban Heng H   Chang Jason S JS   Rubin J Peter JP  

Annals of emergency medicine 20200903 1


<h4>Study objective</h4>Throughout the coronavirus disease 2019 pandemic, many emergency departments have been using passive protective enclosures ("intubation boxes") during intubation. The effectiveness of these enclosures remains uncertain. We sought to quantify their ability to contain aerosols using industry standard test protocols.<h4>Methods</h4>We tested a commercially available passive protective enclosure representing the most common design and compared this with a modified enclosure t  ...[more]

Similar Datasets

| S-BSST1055 | biostudies-other
| 2346925 | ecrin-mdr-crc
| S-EPMC9682111 | biostudies-literature
| S-EPMC8269789 | biostudies-literature
| S-EPMC8548442 | biostudies-literature
| 2383510 | ecrin-mdr-crc
| S-EPMC7267800 | biostudies-literature
| S-EPMC8694760 | biostudies-literature
| S-EPMC7221377 | biostudies-literature
| S-EPMC9124563 | biostudies-literature