Unknown

Dataset Information

0

HbA1c determination from HemaSpot™ blood collection devices: comparison of home prepared dried blood spots with standard venous blood analysis.


ABSTRACT: AIM:To assess the clinical performance and patient acceptance of HemaSpot™ blood collection devices as an alternative blood collection method. METHODS:Adult men and women with any type of diabetes, routinely carrying out self-monitoring of blood glucose were recruited (n = 128). Participants provided a venous blood sample and prepared two HemaSpot dried blood spots, one at clinics and one at home. HbA1c analysis was by Tosoh G8 high-performance liquid chromatography. Participants also completed a questionnaire. RESULTS:Strong linear relationships been HbA1c levels in dried blood spots and venous blood were observed and a linear model was fitted to the data. Time between dried blood spot preparation and testing did not impact the model. Participants were accepting of the approach: 69.2% would use this system if available and 60.7% would be more likely to use this system than going to their general practitioner. CONCLUSIONS:The combination of a robust desiccating dried blood spot device, home sample preparation and return by post produces HbA1c data that support the use of a time-independent linear calibration of dried blood spot to venous blood HbA1c . A robust remote sample collection service would be valuable to people living with diabetes in urban areas who are working or house-bound as well as those living in remote or rural locations.

SUBMITTER: Hall JM 

PROVIDER: S-EPMC7496699 | biostudies-literature | 2020 Sep

REPOSITORIES: biostudies-literature

altmetric image

Publications

HbA<sub>1c</sub> determination from HemaSpot™ blood collection devices: comparison of home prepared dried blood spots with standard venous blood analysis.

Hall J M JM   Fowler C F CF   Barrett F F   Humphry R W RW   Van Drimmelen M M   MacRury S M SM  

Diabetic medicine : a journal of the British Diabetic Association 20191017 9


<h4>Aim</h4>To assess the clinical performance and patient acceptance of HemaSpot™ blood collection devices as an alternative blood collection method.<h4>Methods</h4>Adult men and women with any type of diabetes, routinely carrying out self-monitoring of blood glucose were recruited (n = 128). Participants provided a venous blood sample and prepared two HemaSpot dried blood spots, one at clinics and one at home. HbA<sub>1c</sub> analysis was by Tosoh G8 high-performance liquid chromatography. Pa  ...[more]

Similar Datasets

| S-EPMC5395583 | biostudies-literature
| S-EPMC3698306 | biostudies-literature
| S-EPMC5180185 | biostudies-literature
| S-EPMC3609538 | biostudies-literature